Lumina Power LDD-1000 Series Скачать руководство пользователя страница 11

02001027 
Rev 1 ECO 7062 

Operation Manual: 

LDD-600/1000/1500-XX-YY 

Page 11 of 13 

 

 

 
 
 
 
 
 
 

 

AC input power requirements for LDD-600/1000/1500 models are as follows: 

 

Table 2 

LDD-600/1000/1500 AC Input Power Requirements

 

MODEL INPUT 

POWER 

LDD-600-XX-YY 

100-240 VAC, 50/60 Hz, 7A @115VAC 

LDD-1000-XX-YY 100-240 

VAC, 

50/60 Hz, 12A @115VAC 

LDD-1500-XX-YY 200-240 

VAC, 

50/60 Hz, 9A @220VAC 

 

 

b.

 

INTERFACE SETTINGS

:  Make sure 

INTERLOCK

, Pin 3, is 

connected to GND.  

c.

 

APPLY INPUT AC POWER  

 Turn ON AC power.  After a few seconds 

the power supply fans should begin to run. 

d.

 

PROGRAMMING OUTPUT CURRENT   

Program LDD-

600/1000/1500 power supply for desired output current.  Once the unit has 
been 

ENABLED

 via Pin 1, a 0-10V signal applied to 

Iprogram

, Pin 7, 

will program the LDD-600/1000/1500 diode driver for 0 to maximum rated 
output current.   

 
 
 
 
 
 
 

 
e.

 

ENABLE OUTPUT

   Apply +5V to +15V to 

ENABLE

, Pin 1.  The LDD-

600/1000/1500 will deliver output current as programmed. 

 

5.

 

Monitoring LDD output and performance:

 

a.

 

Current Monitor   

Power supply output current can be monitored via pin 

6, 

Iout Monitor

.  A 0-10V signal will represent the output current from 0 

to maximum rated output current. 

b.

 

Voltage Monitor   

Power supply output voltage can be monitored via pin 

5, 

Vout Monitor

. A 0-10V signal will represent the output voltage from 0-

maximum rated output voltage. If maximum compliance voltage is less 
than 10V, 

Vout Monitor

 will read output voltage directly.  If maximum 

compliance voltage is greater than 10V, then 

Vout Monitor

 will be scaled 

such that 0-10V = 0-Vout

max

.

   

IMPORTANT APPLICATION NOTE 

When the power supply is enabled using the ENABLE signal, internal soft 
start functions limit the rise time of the output current to approximately 
20msec.  Once the power supply is enabled, the rise/fall time of the 
Ip) signal is approximately 600usec.   

IMPORTANT SYSTEM NOTE ON AC INPUT POWER 

 
LDD-600/1000/1500 units are fused on both input lines.  It does not matter 
which of the two AC inputs are designated Line or Neutral.   

Содержание LDD-1000 Series

Страница 1: ...tage capability depending on the user s requirements LDD power supplies will deliver current as programmed into any load providing the voltage requirements of that load do not exceed the maximum rated...

Страница 2: ...s 5 LDD 600 1000 1500 Interface 7 INSTALLATION AND OPERATION 9 Optional RS 232 Protocol 12 Tables and Figures LDD Laser Diode Power Supply Block Diagram 4 LDD 600 1000 1500 Outline Drawing 6 LDD 600 1...

Страница 3: ...er This inverter boosts the input voltage to 400VDC In the process of boosting the input AC voltage the input AC current is adjusted so that is always in phase with the input AC voltage Without this p...

Страница 4: ...e EMI noise and a reduction in overall system heatsink requirements Output Circuit The output filter is a two stage RC filter designed to keep ripple and output noise very low Control Circuit The cont...

Страница 5: ...rent Voltage Monitor 0 10V for 0 Max Voltage Performance Rise Fall Time 600usec 10 to 90 Full Current Current Regulation 0 5 of Maximum output current Temperature Drift 0 5 over temperature range afte...

Страница 6: ...02001027 Rev 1 ECO 7062 Operation Manual LDD 600 1000 1500 XX YY Page 6 of 13 Figure 2 LDD 600 1000 1500 Outline Drawing...

Страница 7: ...y 5 Vout Monitor output 0 10V 0 Voutmax The output voltage of the supply can be monitored by Vout Monitor 6 Iout Monitor output 0 10V 0 Ioutmax The output current of the supply can be monitored by Iou...

Страница 8: ...02001027 Rev 1 ECO 7062 Operation Manual LDD 600 1000 1500 XX YY Page 8 of 13 Figure 3 LDD 600 1000 1500 Interface...

Страница 9: ...re 4 Although the user interface is typically designed by the user Lumina Power can provide any assistance necessary to modify interface program and monitor levels See Table 1 and Figure 3 for descrip...

Страница 10: ...gram Pin 7 A 0 10V signal results in 0 to Ioutmax as long as the rated compliance voltage of the driver is not exceeded 4 Operating the LDD a AC INPUT POWER CONNECTION Connect AC power connections to...

Страница 11: ...ver output current as programmed 5 Monitoring LDD output and performance a Current Monitor Power supply output current can be monitored via pin 6 Iout Monitor A 0 10V signal will represent the output...

Страница 12: ...supply and reading back the output current and voltage The output can be turned on and off with a command as well Command Reply Function Pxx xx xx xx cr cr Program output current to max rated output c...

Страница 13: ...Interface Schematic Pin Pin Name Functional Voltage Level Description 3 Interlock input Open OFF Connect to GND RUN The Interlock function can be connected to external interlock switches such as door...

Отзывы: