MAINTENANCE SCHEDULE

Info: Only qualified persons should service and repair your Lumex LF2020 Easy Lift STS / 

LF2090 Bariatric Easy Lift STS. Regular maintenance of your patient lift is necessary to 

ensure continuing proper and safe operation. Read and observe the following recommended 

maintenance schedule:

Item

Inspect 

initially

Inspect / Adjust

daily

monthly

every 6 

months

Base, Casters, and Foot Paddle

Ensure that base is level (all four wheels touch the floor)

x

x

x

Ensure that bolts and nuts are tight

x

x

x

Ensure that casters swivel and roll freely

x

x

x

Apply White Lithium Grease to caster ball bearings

x

Check rubber parts of casters for deflection

x

x

Check welding joints for rust and cracks

x

Ensure that base opens and closes smoothly

x

x

x

x

Ensure that base locks securely in three positions

x

x

x

Ensure that base locking knob is in place and operating properly

x

x

x

Inspect all base hardware for wear

x

Mast
Ensure that mast is fully engaged and locked into mast sleeve

x

x

Inspect mast for bends or deflections

x

Inspect all mast hardware for wear

x

x

Boom

Inspect boom for bends or deflections

x

x

Ensure that boom is centered between the base legs

x

x

x

Inspect all boom hardware for wear

x

Inspect sling hooks for wear or deflections

x

x

Check connections between boom and mast for improper 

connection, looseness, or wear

x

x

Slings and Sling Hardware
Check slings and straps for wear and damage; discard if worn

x

x

Inspect metal/plastic parts for wear or damage

x

x

Lubrication with White Lithium Grease
Lubricate all pivot points

x

Lubricate top and bottom of actuator mounting

x

Cleaning
Clean with mild non-abrasive detergent

x

Actuator and Electronics
Ensure that batteries are charged

x

x

Inspect for proper up / down operation

x

x

Ensure that mounting hardware is tight

x

x

Inspect mounting hardware for wear

x

Ensure that control box is firmly affixed to the mast

x

x

Ensure that the actuator connecting cable is tight

x

x

Содержание LF2020 Easy Lift STS

Страница 1: ...LF2020 INS LAB RevG18 Read this manual before operating your Easy Lift Sit to Stand Save this manual for future use LF2020 EASY LIFT SIT TO STAND LF2090 BARIATRIC EASY LIFT SIT TO STAND USER MANUAL...

Страница 2: ...TRANSPORTATION INSTRUCTIONS 7 5 ASSEMBLY INSTRUCTIONS 8 CHARGING THE BATTERIES 9 CHECKING BATTERY CAPACITY 9 ALARM BUZZERS 9 6 OPERATING INSTRUCTIONS 10 LIFTING THE PATIENT 11 7 CARE AND MAINTENANCE 1...

Страница 3: ...est that the operator try all the various lifting slings and equipment until the operator is confident with how the patient lift operates The lifting procedures should be explained to the patient befo...

Страница 4: ...te personal injury s NOTICE Indicates a potential hazard situation or unsafe practice that if not avoided could result in product or property damage Info Provides application recommendations or other...

Страница 5: ...s in the sling WARNING Do not attempt to maneuver the lift by pushing on the mast boom actuator or patient Always maneuver the lift with the handle WARNING During lifting or lowering whenever possible...

Страница 6: ...FEATURES EASY LIFT SIT TO STAND FEATURES Pendant Actuator Lifting Boom Knee Pad Footrest Base Brake Mast Control Box CONTROL BOX FEATURES RESET Button Battery Testing Battery Charger Plug In Actuator...

Страница 7: ...s contents If damage is evident notify the carrier or your GF authorized distributor 2 Remove all loose packing from the carton 3 Carefully remove all the components from the carton Info Unless the pa...

Страница 8: ...4 Attach boom to mast and hand tighten bolt and nut 5 Attach the L shaped arms to the back of the knee pad 6 Attach knee pad assembly to the mast 7 Place the control box in the bracket 8 Insert the ac...

Страница 9: ...e control box Frequent plugging and unplugging of the pendant and actuator to and from the control box may damage the control box CHECKING BATTERY CAPACITY Info To check remaining battery capacity pre...

Страница 10: ...ither open or closed position Preparation before lifting 1 Turn on the power by turning the RED RESET BUTTON clockwise 2 Press the pendant UP or DOWN button once 3 After the battery indicator lights d...

Страница 11: ...raps to the hooks 5 Have the patient s hands holding the handles For patients who cannot hold the handles have them hold their arms around the chest 6 Press the UP button on the hand pendant 7 Before...

Страница 12: ...ave them hold their arms around the chest 6 Press the UP button on the hand control 7 Before the patient s body is lifted from the chair stop and make sure the sling is secured and patient s knees are...

Страница 13: ...0 368 4700 for further information When you believe a component or part is not functioning properly immediately contact GF Tech Support at 1 770 368 4700 as a potentially hazardous condition could exi...

Страница 14: ...ting properly x x x Inspect all base hardware for wear x Mast Ensure that mast is fully engaged and locked into mast sleeve x x Inspect mast for bends or deflections x Inspect all mast hardware for we...

Страница 15: ...operate even though battery indicator shows charge Battery charger still plugged in Poor connection between control box and pendant or defective pendant Poor connection between actuator and control bo...

Страница 16: ...gth 50 Lifting height range Info As measured from sling hooks Minimum 27 Maximum 71 Range 35 Mast height 50 Maximum weight capacity LF2020 Easy Lift STS 400 lb EVENLY DISTRIBUTED LF2090 Bariatric Easy...

Страница 17: ...NING WARRANTY SERVICE This limited warranty shall only apply to defects that are reported to the Distributor from whom the Customer purchased the product within the applicable warranty period If there...

Страница 18: ...LOYEE OR INDEPENDENT CONTRACTOR WAGES PAYMENTS AND BENEFITS The warranties contained herein contain all the representations and warranties with respect to the subject matter of this document and super...

Страница 19: ...spection 7 Intended use 3 Intended use contraindication 3 Introduction 3 L Lift and transfer from bed 11 Lift and transfer from chair 12 Lifting the patient 11 Lubrication 13 M Maintenance schedule 14...

Страница 20: ...can be found on our website www grahamfield com 2005 GF Health Products Inc All Rights Reserved Graham Field and Lumex are trademarks of GF Health Products Inc One Graham Field Way Atlanta GA 30340 31...

Отзывы: