8  TROUBLESHOOTING

The following list of encountered problems and their solutions will assist you in determining 
what may be causing your Lumex lift not to function as designed. If you have a problem occurring 
which is not listed below, contact GF Technical Support at 1.770.368.4700 for help. Do not attempt 
to repair components or parts on your lift, as this may invalidate your warranty or cause further 
problems that may result in patient injury. Stop using your lift immediately if it is not functioning 
correctly or any warning beeps are heard.

SYMPTOM

POSSIBLE CAUSE

ACTION

The actuator is stalled and a 

clicking sound is audible

Actuator plug is not seated 

properly into control box

Actuator is defective or at end of 

its life cycle

Control box is defective

Firmly plug actuator plug into 

control box

Replace actuator if control box is 

good

Replace control box if bad; replace 

control box if pendant is good; 

replace pendant if bad

The actuator will either go up or 

down, but not both

Hand pendant or control box 

defective

Replace control box if pendant is 

good

Replace pendant if it is bad

The actuator makes a grinding 

sound while lifting, but works fine 

without a load

The internal gear is stripped, likely 

due to overloading

Stop using the lift immediately!

Repair or replace the actuator

The actuator stops and starts 

during lifting or lowering

Bad batteries

Check battery indicator; confirm 

if lights drop quickly to red light; 

replace batteries

The lift will not operate even though 

battery indicator shows charge

Battery charger still plugged in

Poor connection between control 

box and pendant, or defective 

pendant

Poor connection between actuator 

and control box

Unplug battery charger

Plug hand pendant firmly into 

control box, or replace pendant

Plug actuator cord firmly into 

control box

The lift will not operate and battery 

indicator has no lights illuminated

Control box may be in OFF position

Internally, batteries may be loose 

from their connections

Batteries are in need of being 

charged

Batteries are dead and won’t hold a 

new charge

Turn the red emergency stop button 

until it pops up to the ON position

Check inside control box to ensure 

batteries are connected

Use charger to charge batteries, 

ensuring the red emergency stop 

button is in ON or UP position

Replace the batteries

During charging, the battery 

indicator is not illuminated and 

won’t charge

The red emergency stop button is 

in OFF or DOWN position

The battery charger is defective or 

at end of its life cycle

Turn the red emergency stop button 

until it pops up to the ON position

Inspect the battery charger for 

damage and replace if required

The battery indicator indicates need 

for frequent charging and/or lift 

loses power

The batteries are at end of life cycle Replace the batteries

The lift will not operate, and there 

is no clicking sound or light when 

pendant is pressed

The pendant is loose/defective or 

at end of its life cycle

Check battery indicator; replace 

pendant if battery indicator shows 

full charge

  WARNING: If you experience a problem with your lift and are unable to solve it yourself, 

contact GF Tech Support at 1.770.368.4700 or your GF authorized distributor.

Содержание LF2020 Easy Lift STS

Страница 1: ...LF2020 INS LAB RevG18 Read this manual before operating your Easy Lift Sit to Stand Save this manual for future use LF2020 EASY LIFT SIT TO STAND LF2090 BARIATRIC EASY LIFT SIT TO STAND USER MANUAL...

Страница 2: ...TRANSPORTATION INSTRUCTIONS 7 5 ASSEMBLY INSTRUCTIONS 8 CHARGING THE BATTERIES 9 CHECKING BATTERY CAPACITY 9 ALARM BUZZERS 9 6 OPERATING INSTRUCTIONS 10 LIFTING THE PATIENT 11 7 CARE AND MAINTENANCE 1...

Страница 3: ...est that the operator try all the various lifting slings and equipment until the operator is confident with how the patient lift operates The lifting procedures should be explained to the patient befo...

Страница 4: ...te personal injury s NOTICE Indicates a potential hazard situation or unsafe practice that if not avoided could result in product or property damage Info Provides application recommendations or other...

Страница 5: ...s in the sling WARNING Do not attempt to maneuver the lift by pushing on the mast boom actuator or patient Always maneuver the lift with the handle WARNING During lifting or lowering whenever possible...

Страница 6: ...FEATURES EASY LIFT SIT TO STAND FEATURES Pendant Actuator Lifting Boom Knee Pad Footrest Base Brake Mast Control Box CONTROL BOX FEATURES RESET Button Battery Testing Battery Charger Plug In Actuator...

Страница 7: ...s contents If damage is evident notify the carrier or your GF authorized distributor 2 Remove all loose packing from the carton 3 Carefully remove all the components from the carton Info Unless the pa...

Страница 8: ...4 Attach boom to mast and hand tighten bolt and nut 5 Attach the L shaped arms to the back of the knee pad 6 Attach knee pad assembly to the mast 7 Place the control box in the bracket 8 Insert the ac...

Страница 9: ...e control box Frequent plugging and unplugging of the pendant and actuator to and from the control box may damage the control box CHECKING BATTERY CAPACITY Info To check remaining battery capacity pre...

Страница 10: ...ither open or closed position Preparation before lifting 1 Turn on the power by turning the RED RESET BUTTON clockwise 2 Press the pendant UP or DOWN button once 3 After the battery indicator lights d...

Страница 11: ...raps to the hooks 5 Have the patient s hands holding the handles For patients who cannot hold the handles have them hold their arms around the chest 6 Press the UP button on the hand pendant 7 Before...

Страница 12: ...ave them hold their arms around the chest 6 Press the UP button on the hand control 7 Before the patient s body is lifted from the chair stop and make sure the sling is secured and patient s knees are...

Страница 13: ...0 368 4700 for further information When you believe a component or part is not functioning properly immediately contact GF Tech Support at 1 770 368 4700 as a potentially hazardous condition could exi...

Страница 14: ...ting properly x x x Inspect all base hardware for wear x Mast Ensure that mast is fully engaged and locked into mast sleeve x x Inspect mast for bends or deflections x Inspect all mast hardware for we...

Страница 15: ...operate even though battery indicator shows charge Battery charger still plugged in Poor connection between control box and pendant or defective pendant Poor connection between actuator and control bo...

Страница 16: ...gth 50 Lifting height range Info As measured from sling hooks Minimum 27 Maximum 71 Range 35 Mast height 50 Maximum weight capacity LF2020 Easy Lift STS 400 lb EVENLY DISTRIBUTED LF2090 Bariatric Easy...

Страница 17: ...NING WARRANTY SERVICE This limited warranty shall only apply to defects that are reported to the Distributor from whom the Customer purchased the product within the applicable warranty period If there...

Страница 18: ...LOYEE OR INDEPENDENT CONTRACTOR WAGES PAYMENTS AND BENEFITS The warranties contained herein contain all the representations and warranties with respect to the subject matter of this document and super...

Страница 19: ...spection 7 Intended use 3 Intended use contraindication 3 Introduction 3 L Lift and transfer from bed 11 Lift and transfer from chair 12 Lifting the patient 11 Lubrication 13 M Maintenance schedule 14...

Страница 20: ...can be found on our website www grahamfield com 2005 GF Health Products Inc All Rights Reserved Graham Field and Lumex are trademarks of GF Health Products Inc One Graham Field Way Atlanta GA 30340 31...

Отзывы: