565G/565PG/565WG Ortho-Biotic Preferred Care
®
Recliners Operating Instructions
8
Figure 13
RECLINER POSITIONING MECHANISM
TROUBLE-SHOOTING
ADJUSTMENT INSTRUCTIONS FOR POSITION
MECHANISM
NOTE: Never lubricate the position
mechanism assembly.
1. Turn the position adjustment handle down to
the release position, and remove clevis pin and
wire ring from top of assembly (See fig. 13).
2. Slide positioning plate up, and turn square
washer clockwise until snug (See fig. 13).
3. Slide the positioning plate down over square
washer. Replace clevis pin and wire ring.
4. The position mechanism's handle should
engage at about the halfway point. If more
tension is desired, repeat the above procedure.
Clevis ring and
wire ring
Positioning plate
Square washer