2.2.4
1409 Multipoint Sampler Configuration:
The
Read Configuration
button show the numbers of Valve blocks mounted in the 1409. It
also makes it possible manually to activate the valves of the 1409.
2.2.5
Testing the Sample valves
Each Sample valve can be activated or deactivated by clicking on the corresponding number in
the BZ7006.
The Analyzer 3-way valves can be activated end deactivated by clicking on the
Analyzer
button
The Waste air outlet 3-way valve can be activated or deactivated by clicking on the
Waste Air
button
The
Valve Sequence
button will start a single test of the valves, including the Analyzer and
Waste Air Valves.
The valve sequence test will continue if the
Repeat Status and Sequence
is marked before
activating the
Valve sequence
.
Stop the test by removing the mark of the
Repeat Status and Sequence.