Chapter 2
____________________________________________________________________
_______________________________________________________________________________________
BI7717-11
1409 Multipoint Sampler
LumaSense Technologies A/S
Service Manual
Page 14 of 28
2.2 Testing procedure
2.2.1
Starting the program
Double click on the BZ7006 icon on your desktop, to start the BZ7006 Service software.
The following buttons and functions are available
“USB Connect”
Connects the attached 1409 to the program
“Reset Command”
Resets the 1409 and close all open valves
“Valve Sequence”
start an automatic test of the sample valves
“Read Configuration”
Reads the numbers of valve blocks connected in the 1409
“Read ID”
Reads out the detailed Identification of the 1409
“*IDN?”
Reads out the Identity of the 1409 in short form
“Read Status”
Reads the status flag from the 1409
“Clear Status”
Clear the status flags in the 1409
2.2.2
Connection of the 1409
Connect the 1409 Multipoint Sampler directly to the computer using a USB cable, INNOVA Part
NO. AS0001B
Now click on the
USB Connect
button, to connect to the attached 1409.