background image

Safety Consideration 

____________________________________________________________________ 

_______________________________________________________________________________________ 

BI7717-11 

1409 Multipoint Sampler 

LumaSense Technologies A/S 

Service Manual  

 

Page 5 of 28 

 

 

SAFETY CONSIDERATIONS 

 

PLEASE 

READ 

THESE 

SAFETY 

CONSIDERATIONS 

CAREFULLY  AND  MAKE  SURE  YOU  UNDERSTAND  THEM 
PROPERLY  BEFORE  YOU  START  OPERATING  THE  1409 
MULTIPOINT SAMPLER. 

 
EXPLOSION HAZARD 

 

THE 1409 MULTIPOINT SAMPLER IS NOT DESIGNED 
FOR USE IN POTENTIALLY EXPLOSIVE 
ENVIRONMENTS. 

 

 

This means that the instrument must 

not

 be placed and operated in 

an area with a potentially explosive atmosphere. 

 
 

When  monitoring  potentially  flammable  or  toxic  gases  it  is  essential 
that: 
1)

 

the instrument itself is placed in a well-ventilated area 

outside 

the 

potentially hazardous zone; and 

2)

 

a  sufficiently  long  tube  is  connected  to  the  “Sampler  Waste  Air 
Outlet” so that the sampled gas is carried 

away

 to the open air or 

to an extraction and/or filtration unit. 

 

TO  AVOID  THE  POSSIBILITY  OF  AN  EXPLOSION,  MONITORING  OF 
FLAMMABLE  GASES 

IN  EXPLOSIVE  CONCENTRATIONS 

MUST 

NEVER BE ATTEMPTED. 

 

AVOID WATER CONDENSATION IN THE INSTRUMENT. 

 

Liquids must be prevented from entering the instrument. It is therefore 
important that warm humid gases be not drawn into a cold instrument 
because  condensation  will  take  place.  If  such  a  situation  is  likely  to 
occur, you should ensure that the gases are drawn through water-trap 
filters  before  they  enter  the  sampler  channels  of  the  1409.  This  will 
remove  water  vapour  in  the  gases  and  thus  prevent  condensation 
within  the  instrument.  The  water-trap  filter  should  be  used  in  the 
immediate  environment  of  the  instrument  so  it  maintains  either  the 
same temperature, or a lower temperature than the instrument. 

 

Note: 

that some gases may be absorbed by the water trapped in the 

filter. This will reduce the gases’ concentration. 

 

 
 
 

 
 
 
 

 
 
 

 
 

Содержание Innova-1409

Страница 1: ...__ BI7717 11 Service Manual Multipoint Sampler Innova 1409...

Страница 2: ..._______________________________________________________________________________________ BI7717 11 1409 Multipoint Sampler LumaSense Technologies A S Service Manual Page 2 of 28...

Страница 3: ...System requirements 13 2 1 2 Installation of the program 13 2 2 Testing procedure 14 2 2 1 Starting the program 14 2 2 2 Connection of the 1409 14 2 2 3 1409 Multipoint Sampler Id identification 15 2...

Страница 4: ...________________________________________ _______________________________________________________________________________________ BI7717 11 1409 Multipoint Sampler LumaSense Technologies A S Service Ma...

Страница 5: ...iciently long tube is connected to the Sampler Waste Air Outlet so that the sampled gas is carried away to the open air or to an extraction and or filtration unit TO AVOID THE POSSIBILITY OF AN EXPLOS...

Страница 6: ...__________________________________ _______________________________________________________________________________________ BI7717 11 1409 Multipoint Sampler LumaSense Technologies A S Service Manual P...

Страница 7: ...e available for field service Handling of the electronic parts must be done with proper ESD equipment to avoid any possibility of an electrostatic discharge to the components Modules returned to LumaS...

Страница 8: ..._______________________ _______________________________________________________________________________________ BI7717 11 1409 Multipoint Sampler LumaSense Technologies A S Service Manual Page 8 of 28...

Страница 9: ...Technologies A S Service Manual Page 9 of 28 1 2 2 Front Panel Fig 1 2 The front panel of the 1409 LED indicators Each channel is numbered and has a correspondingly numbered lamp When the lamp is lit...

Страница 10: ...the corresponding sampling valve is open AC Mains A 3 pin connector accepting Power Cable for connection to a single phase AC mains supply with protective Earth Mains Voltage Connect 1409 to mains sup...

Страница 11: ...___________________________________ _______________________________________________________________________________________ BI7717 11 1409 Multipoint Sampler LumaSense Technologies A S Service Manual...

Страница 12: ...____________________________ _______________________________________________________________________________________ BI7717 11 1409 Multipoint Sampler LumaSense Technologies A S Service Manual Page 12...

Страница 13: ...g a Microsoft Windows Operating System Before installing the Software the PC must meet the following minimum requirements BZ7006 Software Computer requirements Processor Minimum 1 gigahertz GHz Pentiu...

Страница 14: ...onnect Connects the attached 1409 to the program Reset Command Resets the 1409 and close all open valves Valve Sequence start an automatic test of the sample valves Read Configuration Reads the number...

Страница 15: ...8 2 2 3 1409 Multipoint Sampler Id identification The Read ID button will give you information regarding the 1409 type and firmware configuration Type is 1409 3 way is 1 when the Center Manifold is co...

Страница 16: ...of the 1409 2 2 5 Testing the Sample valves Each Sample valve can be activated or deactivated by clicking on the corresponding number in the BZ7006 The Analyzer 3 way valves can be activated end deac...

Страница 17: ...the firmware 2 2 7 1409 Multipoint Sampler Status The Read Status button will give a read out of the status error log and 12V power status The Clear Status button will reset the error log data Reset i...

Страница 18: ...__________________________ _______________________________________________________________________________________ BI7717 11 1409 Multipoint Sampler LumaSense Technologies A S Service Manual Page 18 o...

Страница 19: ..._______ _______________________________________________________________________________________ BI7717 11 1409 Multipoint Sampler LumaSense Technologies A S Service Manual Page 19 of 28 3 1 Mechanical...

Страница 20: ..._____________________________ _______________________________________________________________________________________ BI7717 11 1409 Multipoint Sampler LumaSense Technologies A S Service Manual Page 2...

Страница 21: ..._____________________________ BI7717 11 1409 Multipoint Sampler LumaSense Technologies A S Service Manual Page 21 of 28 3 1 4 Multipoint Sampler PCB ZD0900A Connector Connection J4 Sample manifold Ch...

Страница 22: ...______________________________ _______________________________________________________________________________________ BI7717 11 1409 Multipoint Sampler LumaSense Technologies A S Service Manual Page...

Страница 23: ..._______________________ BI7717 11 1409 Multipoint Sampler LumaSense Technologies A S Service Manual Page 23 of 28 4 1 Parts replacements Follow these instructions to disassembly the cabinet and gain a...

Страница 24: ..._________________ _______________________________________________________________________________________ BI7717 11 1409 Multipoint Sampler LumaSense Technologies A S Service Manual Page 24 of 28 Remo...

Страница 25: ..._ _______________________________________________________________________________________ BI7717 11 1409 Multipoint Sampler LumaSense Technologies A S Service Manual Page 25 of 28 Remove the Left pane...

Страница 26: ...Description AQ1720A Power Cable DB2974 Tube fitting DD0605A Cover cap for Torx T10 screw DF0238A Decoration strips kit with 4 pcs DF7076A Feet for 1409 kit with 4 pcs DS0894A Filter for Air inlet YJ07...

Страница 27: ..._________________________________ _______________________________________________________________________________________ BI7717 11 1409 Multipoint Sampler LumaSense Technologies A S Service Manual Pa...

Страница 28: ...______________________ __ LumaSense Technologies A S Tel 45 44 20 01 00 Energivej 30 Fax 45 44 20 01 01 DK 2750 Ballerup Denmark http www lumasenseinc com _____________________________________________...

Отзывы:

Похожие инструкции для Innova-1409