IMPAC Infrared sensor IN 2000

 

 

 

 

17

12

 

Data format UPP

 

(

U

niversal 

P

yrometer 

P

rotocol)

 

 

Via interface and suitable communication software or via “Test” function of the 

InfraWin

 software (see 

8

,

 

Basic settings 

 Test

) commands can be exchanged directly with the pyrometer. 

The data exchange occurs in ASCII format with the following transmission parameters: 
The data format is:    8 data bits, 1 stop bit, even parity (8,1,e) 
The device responds to the entry of a command with: output (e.g. the measuring value) + CR (

C

arriage 

R

eturn, ASCII 13), to pure entry commands with “ok” + CR. 

Every command starts with the 2-digit device address AA (e.g. “00”).

 

This is followed by 2 small command 

letters (e.g. “em” for level of emissivity 

), finished with CR 

This is followed, if necessary for that command, by the ASCII parameter “X”. If this parameter “X” is omitted, 
then the device resets with the current parameter. 
A “?“ after the small command letters answers with the respective settings (only at setting commands, not at 
enquiry commands). 

Example

:  Entry: “00em“ + 

CR

 

 

The emissivity setting (

ε

) of the device with the address 00 is returned 

 

Answer: “0970“ + 

CR

 means Emissivity = 0.97 or 97.0% 

 

Description 

Command 

Parameters 

Reading temperature 
value: 

AAms 

Output: XXXXX (in 1/10 °C oder °F) 

(88880 = Temperature overflow) 

Reading temperature 
value repeated: 

AAmsXXX 

XXX = 001...999 (XXX = number of measuring values) 

Emissivity 

ε

AAemXXXX 

XXXX = (0010 ... 1000‰) (decimal) 

Exposure time t

90

AaezX 

X = 0 ... 9 (decimal) 

0 = intrinsic time constant of the device  

 

1 = 0,50 s 

4 = 5,00 s 

7 = 60,00 s 

 

 

2 = 1,00 s 

5 = 10,00 s 

8 = 90,00 s 

 

 

3 = 2,00 s 

6 = 30,00 s 

9 = 120,00 s 

 

Clear time maximum 
value storage: 

AAlzX 

X = 0 ... 8 (dec.) 

0 = Maximum value storage off  

1 = 0.1 s 

4 = 1.00 s 

7 = (not available) 

2 = 0.25 s 

5 =  5.00 s  

8 = automatically deletion 

3 = 0.5 s 

6 = 25.00 s 

Reading basic tem-
perature range: 

AAmb 

Output: XXXXYYYY (hex 8-digit, °C) 

XXXX = beginning of temperature range 
YYYY = end of temperature range 

Reading temperature  
sub range: 

AAme 

Output: XXXXYYYY (hex 8-digit, °C) 

XXXX = beginning of temperature range 
YYYY = end of temperature range 

Setting of temperature  
sub range: 

AAm1XXXXYYYY  XXXX (hex 4-digit) beginning of temp. range (°C) 

YYYY (hex 4-digit) end of temp. range (°C) 

Address: 

AAgaXX 

XX = (00 ... 97)  00 ... 97 = adjustable addresses 

Baud rate: 

AAbrX 

X = 3...4  
 

3 = 9600 Baud  4 = 19200 Baud 

Changing °C / °F 

AAfhX 

Output: X = 0: display in °C;  X = 1: display in °F 

Internal temperature: 

AAgt 

Output: XX (decimal 00 … 98, in °C) 
 

XXX (decimal 032 … 208°F) 

Max. internal  
temperature: 

AAtm 

Output: XX (decimal 00 … 98, in °C) 
 

XXX (decimal 032 … 208°F) 

Error status: 

AAfs 

Output 1 byte hex (00 = no error) 

Reading parameters: 

AApa 

Output decimal 11-digit: 

Digit 1 und 2 (10...99 or 00): Emissivity 
Digit 3 (0 ... 9): Exposure time 
Digit 4 (0 ... 8): Clear time max. storage 
Digit 5 (1): Analog output 
Digit 6 and 7: (00 ... 98): Internal temperature 
Digit 8 and 9 (00 ... 97): Address 
Digit 10 (3 … 4): Baud rate 
Digit 11 (0): always 0 

Device type: 

AAna 

Output: “IN 2000” 

Serial number: 

AAsn 

Output: XXXX (hex 4-digit) 

Device type /  
software version: 

AAve 

Output: XXYYZZ (6-digit decimal) 

XX = 77 (IN 2000) 
YY = Month of software version  
ZZ = Year of software version 

Содержание IN 2000

Страница 1: ...IN 2000 Infrarot Messumformer Infrared Sensor Operation Manual Betriebsanleitung Proven Quality www imaxsun com...

Страница 2: ...maximum value storage tClear 10 7 8 Analog output mA 10 7 9 Subrange from to 10 7 10 Ambient temperature compensation 10 7 11 Address 11 7 12 Operating hours 11 7 13 Baud rate Baud 11 8 Software Infr...

Страница 3: ...e right to revise this document and to make changes from time to time in the content hereof without obligation to notify any person or persons of such revisions or changes LumaSense Technologies offer...

Страница 4: ...s for max min value storage or automatically deletion of the storage temperature sub range ambient temperature compensation address baud rate Maximum minimum value storage Clearing with adjusted time...

Страница 5: ...re measuring curves 2 2 Scope of delivery IN 2000 inspection sheet operation manual 3 Safety 3 1 General Each person working with the pyrometer must have read and understood the user manual before ope...

Страница 6: ...al output yellow TxD transmit data of digital output To meet the electromagnetic requirements EMV a shielded connecting cable must be used The shield of the connecting cable has to be connected only o...

Страница 7: ...sealing screw 5 1 Accessories option The overview on the right side shows a selection of suitable accessories You can find the entire accessory program with all reference numbers on section 13 Refer e...

Страница 8: ...ameter descriptions settings The pyrometer is equipped with a wide range of settings for optimal adaptation to the required measuring condition and to measure the temperature correct All settings can...

Страница 9: ...listed below The tolerance of the emissivity values for each material is mainly dependent on the surface conditions Rough surfaces have higher emissivities Measuring object at 8 14 m Measuring object...

Страница 10: ...justment of exposure time This results in clear time the adjusted response time is useless clear times must be at least 3 times longer than the response time only maxima with full maximum value can be...

Страница 11: ...utilizing the software functions the USB adapter is required accessory Additionally to parameter adjustments the software InfraWin features the possibility of recording measure ment values inclusive...

Страница 12: ...terface connected to the pyrometer has to be selected under the Computer icon For two devices two PC interfaces must be used 8 6 Number of devices With a click on number of devices InfraWin changes to...

Страница 13: ...f the temperature range or the adjusted sub range Entering temperature values in the white fields on the left and right side of the color bar limits for the color change of the color bar can be set Th...

Страница 14: ...alues of measurement color bar or measurement online trend are automatically saved as standard i12 Should you need to edit the data later you need to save the file as another i12 file because old valu...

Страница 15: ...andard i12 file and automatically appears in this form upon opening Listing or Trend output If file open was loaded using another file the previous file will be overwritten and re placed by the standa...

Страница 16: ...or and to solve the problem with the help of the following list Temperature indication too low Incorrect alignment of the pyrometer to the object New correct alignment to achieve the max temperature s...

Страница 17: ...t90 AaezX X 0 9 decimal 0 intrinsic time constant of the device 1 0 50 s 4 5 00 s 7 60 00 s 2 1 00 s 5 10 00 s 8 90 00 s 3 2 00 s 6 30 00 s 9 120 00 s Clear time maximum value storage AAlzX X 0 8 dec...

Страница 18: ...ents 3 885 200 IN 2000 13 2 Reference numbers accessories 3 890 600 Power supply 24 V DC 3 826 650 USB Adapter 3 837 180 Cooled enclosure 3 835 250 90 mirror 3 834 260 Mounting angle fixed 3 834 250 M...

Отзывы: