IMPAC Infrared sensor IN 2000

 

 

 

 

10 

7.7 

Clear time of the maximum value storage (tClear) 

If the maximum value storage is switched on always the highest last temperature value 
will be displayed and stored. The storage has to be cleared at regular intervals for 
exchanging by a new and actual value.  
This feature is particularly useful when fluctuating object temperatures cause the dis-
play or the analog outputs to change too rapidly, or the pyrometer is not constantly 
viewing an object to be measured. In addition, it may also be beneficial to periodically 
delete and reset the stored maximum values.  
 
The following settings are possible: 

off: 

At clear time “off“ the max. value storage is switched off and only momentary values are meas-
ured. 

0.01...25 s: 

If any clear time between 0.01 s and 25 s is set, the maximum value is estimated and held in 

double storage mode.

 After the entered time the storage will be deleted. 

auto: 

The “

auto

” mode is used for discontinuous measuring tasks. For example objects are trans-

ported on a conveyer belt and pass the measuring beam of the pyrometer only for a few sec-
onds. Here the maximum value for each object has to be indicated. In this mode the maximum 
value is stored until a new hot object appears in the measuring beam. The temperature which 
has to be recognized as “hot“ is defined by the low limit of the adjusted sub range. The stored 
maximum value will be deleted when the temperature of the new hot object exceeds the low 
limit “

from

“ of the sub range by 1% or at least 2°C. If a lower limit is not entered, the maximum 

value storage will be deleted whenever the lower level of the full measuring range has been ex-
ceeded.  

 

 

Note:  

The maximum value storage follows the function of adjustment of exposure time.  
This results in: 



clear time 

 the adjusted response time is useless 

 

 



clear times must be at least 3 times longer than the response time 



only maxima with full maximum value can be recorded, which 

 

   appear at least 3 times longer than the response time. 

 
 

7.8 

Analog output (mA) 

The analog output has to be selected according to the signal input of the connected 
instrument (controller, PLC, etc.). 
 
 

7.9 

Subrange (from / to) 

You have the opportunity to choose a subrange (minimum 51°C) within the basic temperature range of the 
pyrometer. This subrange corresponds to the analog output. “

from

“ describes the beginning of this tempera-

ture range, “

to

“ the end of the range.  

Additionally with the setting of a subrange it is possible to fulfill the requirements of the “auto” clear mode of 
the maximum value storage (see above). 
 
 

7.10 

Ambient temperature compensation  

The compensation of the ambient temperature can be used at low measuring tempera-
tures (below 100°C). This compensation is used for a very few special applications only. 
The standard setting of this parameter is “auto”, because the temperature of the air 
around the pyrometer is normally the ambient temperature of the measured object. 
Should the measured object be placed in an area with a higher wall temperature (e.g. 
inside a furnace), the measurement might be falsified (probably too high temperature indication). This influ-
ence can be compensated by presetting of the ambient temperature of the object (presetting within the 
measuring range of the instrument).  
After switching over to “man” the corresponding data field T(amb) is activated so that the ambient tempera-
ture value can be entered.  

 

 

 

It has to be considered, that this method only improves the results if the ambient temperature 

at the place of the measured object is always constant and the emissivity is well known. 

 

Settings: 

off 

0.1 s 

. . . 

25 s 
auto 

Set to: 

4 ... 20 mA 

Settings: 

auto 

-32°C (-25,6°F) 

. . . 

900°C (1652°F) 

Содержание IN 2000

Страница 1: ...IN 2000 Infrarot Messumformer Infrared Sensor Operation Manual Betriebsanleitung Proven Quality www imaxsun com...

Страница 2: ...maximum value storage tClear 10 7 8 Analog output mA 10 7 9 Subrange from to 10 7 10 Ambient temperature compensation 10 7 11 Address 11 7 12 Operating hours 11 7 13 Baud rate Baud 11 8 Software Infr...

Страница 3: ...e right to revise this document and to make changes from time to time in the content hereof without obligation to notify any person or persons of such revisions or changes LumaSense Technologies offer...

Страница 4: ...s for max min value storage or automatically deletion of the storage temperature sub range ambient temperature compensation address baud rate Maximum minimum value storage Clearing with adjusted time...

Страница 5: ...re measuring curves 2 2 Scope of delivery IN 2000 inspection sheet operation manual 3 Safety 3 1 General Each person working with the pyrometer must have read and understood the user manual before ope...

Страница 6: ...al output yellow TxD transmit data of digital output To meet the electromagnetic requirements EMV a shielded connecting cable must be used The shield of the connecting cable has to be connected only o...

Страница 7: ...sealing screw 5 1 Accessories option The overview on the right side shows a selection of suitable accessories You can find the entire accessory program with all reference numbers on section 13 Refer e...

Страница 8: ...ameter descriptions settings The pyrometer is equipped with a wide range of settings for optimal adaptation to the required measuring condition and to measure the temperature correct All settings can...

Страница 9: ...listed below The tolerance of the emissivity values for each material is mainly dependent on the surface conditions Rough surfaces have higher emissivities Measuring object at 8 14 m Measuring object...

Страница 10: ...justment of exposure time This results in clear time the adjusted response time is useless clear times must be at least 3 times longer than the response time only maxima with full maximum value can be...

Страница 11: ...utilizing the software functions the USB adapter is required accessory Additionally to parameter adjustments the software InfraWin features the possibility of recording measure ment values inclusive...

Страница 12: ...terface connected to the pyrometer has to be selected under the Computer icon For two devices two PC interfaces must be used 8 6 Number of devices With a click on number of devices InfraWin changes to...

Страница 13: ...f the temperature range or the adjusted sub range Entering temperature values in the white fields on the left and right side of the color bar limits for the color change of the color bar can be set Th...

Страница 14: ...alues of measurement color bar or measurement online trend are automatically saved as standard i12 Should you need to edit the data later you need to save the file as another i12 file because old valu...

Страница 15: ...andard i12 file and automatically appears in this form upon opening Listing or Trend output If file open was loaded using another file the previous file will be overwritten and re placed by the standa...

Страница 16: ...or and to solve the problem with the help of the following list Temperature indication too low Incorrect alignment of the pyrometer to the object New correct alignment to achieve the max temperature s...

Страница 17: ...t90 AaezX X 0 9 decimal 0 intrinsic time constant of the device 1 0 50 s 4 5 00 s 7 60 00 s 2 1 00 s 5 10 00 s 8 90 00 s 3 2 00 s 6 30 00 s 9 120 00 s Clear time maximum value storage AAlzX X 0 8 dec...

Страница 18: ...ents 3 885 200 IN 2000 13 2 Reference numbers accessories 3 890 600 Power supply 24 V DC 3 826 650 USB Adapter 3 837 180 Cooled enclosure 3 835 250 90 mirror 3 834 260 Mounting angle fixed 3 834 250 M...

Отзывы: