background image

IMPAC pyrometers IS 210 · IGA 210

 

 

 

 

12 

10.2 

Cleaning 

The pyrometer has no internal parts, which have to be serviced. The lens can be cleaned with 
compressed air, which is dry and free of oil. If the lens requires more thorough cleaning, use a 
soft, dry cloth such as that used to clean camera lenses. 

Caution:  Do not clean or touch the lens with acids or solvents!

 

 
 

11

 

Trouble shooting 

 
Before sending the pyrometer for repair, try to find the error and to solve the problem with the 
help of the following list. 

Temperature indication too low 

  Incorrect alignment of the pyrometer to the object 

 New correct alignment to achieve the max. temperature signal 

  Measuring object smaller than spot size 

 check measuring distance, smallest spot size is at nominal measuring distance (see 

5

  Emissivity set too high 

 Set lower correct emissivity corresponding to the material (see 

7.2

 Lens 

contaminated 

 Clean lens carefully (see 

10.2

Temperature indication too high 

  Emissivity set too low 

 Set higher correct emissivity corresponding to the material (see 

7.2

  The measurement is influenced by reflections of hot machine parts 

 Use mechanical construction to avoid the influence of the interfering radiation 

Measuring errors 

  Indicated temperature is decreasing during the use of the pyrometer, contamination of the lens 

 Clean lens. Recommendation: use of air purge (see 

4

  Indicated temperature is decreasing during the use of the pyrometer, although the air 

purge unit is used. Probably compressed air is not clean or air failed 

 Clean the lens and use clean, dry and oil free compressed air 

 HF-interferences 

 

 Correct the connection of the cable shield (see 

3

 Instrument 

overheated 

 Use cooling jacket with air or water cooling 

 
 

Содержание IGA 210

Страница 1: ...IS 210 IGA 210 IMPAC Pyrometer Proven Quality Operation Manual Betriebsanleitung...

Страница 2: ...a USB adapter onto a PC 6 3 4 Connection to the portable parametrizing device HT 6000 7 4 Mechanical Installation accessories 7 5 Optics 7 6 Alignment with the LED targeting light 8 7 Parameter Settin...

Страница 3: ...s that arise from the use of any exam ples or processes mentioned in this manual or in case the content of this document should be incomplete or incorrect LumaSense Technologies reserves the right to...

Страница 4: ...peratures 75 C the output signal is set to approx 3 9 mA Load max 700 at 24 V power supply 400 18 V 1000 30 V Service interface for parameter programming with portable setup device HT 6000 or via USB...

Страница 5: ...dditional devices operating with this mains voltage e g transformers Touching mains voltage can be fatal An incorrect connection and mounting can cause serious health or material damages Only qualifie...

Страница 6: ...corder Digital display 230 V 4 20 mA white brown Schirm schwarz 24 V DC Power supply green yellow Power supply for LED targeting light White 24 V Pin 1 Brown 0 V Pin 2 Black Shield Green LED Pin 5 Yel...

Страница 7: ...least as big as the spot size of the pyrometer in the meas uring distance The following drawings show the pyrometer s spot sizes M in different measuring distances a The spot size M in a distance of a...

Страница 8: ...ch are different from the nominal distance of the selected optics may be calculated using the following formula D D M a a M 2 2 D D M a a M 1 1 6 Alignment with the LED targeting light For exact measu...

Страница 9: ...r the two spectral ranges of the in struments are listed in the emissivity table below The tolerance of the emissivity values for each material is mainly dependent on the surface conditions Rough surf...

Страница 10: ...until a new hot object appears in the measuring beam The temperature which has to be recognized as hot is defined by the low limit of the adjusted sub range The stored maximum value will be deleted wh...

Страница 11: ...surement curves Subsequent analysis graphically or in table form for print out or export Spot size calculator to calculate spot sizes in any distances 9 Transport packaging storage With faulty shippin...

Страница 12: ...suring distance see 5 Emissivity set too high Set lower correct emissivity corresponding to the material see 7 2 Lens contaminated Clean lens carefully see 10 2 Temperature indication too high Emissiv...

Страница 13: ...60 Connection cable 5 m 3 821 770 Connection cable 10 m 3 821 780 Connection cable 15 m 3 821 790 Connection cable 20 m 3 821 800 Connection cable 25 m 3 821 810 Connection cable 30 m 3 826 500 Portab...

Отзывы: