background image

IMPAC pyrometers IS 210 · IGA 210

 

 

 

 

1.2 

Appropriate use 

The instruments IS 210 and IGA 210 are digital pyrometers for non-contact temperature meas-
urement. They are generally used for measurements of metals, ceramics or graphite. 
The infrared radiation from the object is imaged to the detector and converted into an electrical 
signal. This signal is linearized and converted into the standard analog current output  
4 to 20 mA. 
 
 

1.3 

Scope of delivery 

Device with selectable optics, 2 fixing nuts SW 46. 
 
 

2

 

Safety 

 

2.1 

General 

Each person working with the pyrometer must have read and understood the user manual 
before operation. Also this has to be done if the person already used similar instruments or 
was already trained by the manufacturer.  
The pyrometer has only to be used for the purpose described in the manual. It is recom-
mended to use only accessories offered by the manufacturer.  
 
 

2.2 

Electrical connection 

Follow common safety regulations for mains voltage (230 or 115 V AC) connecting additional 
devices operating with this mains voltage (e.g. transformers). Touching mains voltage can be 
fatal. An incorrect connection and mounting can cause serious health or material damages.  
Only qualified specialists are allowed to connect such devices to the mains voltage.  
 
 

3

 

Electrical Installation 

 
The instruments of series 210 are supplied with 24 V DC (18 to 30 V DC). When connecting 
the device to the power supply ensure correct polarity. The power consumption (in this case 
4 to 20 mA) is also the measuring signal. The instrument doesn’t need any time for starting or 
preheating and is immediately ready for operation.  
 
IS 210 and IGA 210 are equipped with a LED targeting light for aiming at the measuring ob-
ject. The center of the light marks the center of the spot. For use of this targeting light a sepa-
rate power supply is necessary with 5 to 30 V DC and 35 mA. 
 
To meet the electromagnetic requirements, a shielded connecting cable must be used. The 
shield of the connecting cable is usually only connected on the pyrometer side. If the connect-
ing cable is extended, the extension cable also needs to be shielded. Do not connect the 
shield in the control cabinet to avoid ground loops.  
 

Содержание IGA 210

Страница 1: ...IS 210 IGA 210 IMPAC Pyrometer Proven Quality Operation Manual Betriebsanleitung...

Страница 2: ...a USB adapter onto a PC 6 3 4 Connection to the portable parametrizing device HT 6000 7 4 Mechanical Installation accessories 7 5 Optics 7 6 Alignment with the LED targeting light 8 7 Parameter Settin...

Страница 3: ...s that arise from the use of any exam ples or processes mentioned in this manual or in case the content of this document should be incomplete or incorrect LumaSense Technologies reserves the right to...

Страница 4: ...peratures 75 C the output signal is set to approx 3 9 mA Load max 700 at 24 V power supply 400 18 V 1000 30 V Service interface for parameter programming with portable setup device HT 6000 or via USB...

Страница 5: ...dditional devices operating with this mains voltage e g transformers Touching mains voltage can be fatal An incorrect connection and mounting can cause serious health or material damages Only qualifie...

Страница 6: ...corder Digital display 230 V 4 20 mA white brown Schirm schwarz 24 V DC Power supply green yellow Power supply for LED targeting light White 24 V Pin 1 Brown 0 V Pin 2 Black Shield Green LED Pin 5 Yel...

Страница 7: ...least as big as the spot size of the pyrometer in the meas uring distance The following drawings show the pyrometer s spot sizes M in different measuring distances a The spot size M in a distance of a...

Страница 8: ...ch are different from the nominal distance of the selected optics may be calculated using the following formula D D M a a M 2 2 D D M a a M 1 1 6 Alignment with the LED targeting light For exact measu...

Страница 9: ...r the two spectral ranges of the in struments are listed in the emissivity table below The tolerance of the emissivity values for each material is mainly dependent on the surface conditions Rough surf...

Страница 10: ...until a new hot object appears in the measuring beam The temperature which has to be recognized as hot is defined by the low limit of the adjusted sub range The stored maximum value will be deleted wh...

Страница 11: ...surement curves Subsequent analysis graphically or in table form for print out or export Spot size calculator to calculate spot sizes in any distances 9 Transport packaging storage With faulty shippin...

Страница 12: ...suring distance see 5 Emissivity set too high Set lower correct emissivity corresponding to the material see 7 2 Lens contaminated Clean lens carefully see 10 2 Temperature indication too high Emissiv...

Страница 13: ...60 Connection cable 5 m 3 821 770 Connection cable 10 m 3 821 780 Connection cable 15 m 3 821 790 Connection cable 20 m 3 821 800 Connection cable 25 m 3 821 810 Connection cable 30 m 3 826 500 Portab...

Отзывы: