Model 3001 Multi-Detector Survey Meter
User’s Manual
Section 4
Ludlum Measurements, Inc.
Page 4-12
April 2017
Primary Count Units and Minimum Display (Default 0 µR) -
Tap
ON/ACK to select the value to adjust, and MODE to adjust the value.
Available values are:
Number of Decimal Places (0, 1, or 2)
Minimum Units depend on the selected Primary Units (See list below.)
Primary
Units
Primary
Count
Units
Available
Primary
Units
Primary
Count
Units
Available
cps
kcps
cps
c
µR/h
mR/h
R/h
µR/h µR
mR/h mR
R/h R
cpm
kcpm
cpm
c
Bq
kBq
Bq
d
µSv/h
mSv/h
Sv/h
µSv/h µSv
mSv/h mSv
Sv/h Sv
dpm
kdpm
dpm
d
Primary Count Alarm Point (Default 000) -
Tap ON/ACK to select the
value to adjust, and MODE to adjust the value. Units will be the same as
selected earlier with Primary Count Units. Primary Count Units of c or d will
not be displayed, but mR or µSv will. The ALARM LCD Segment will be on to
indicate an Alarm parameter. Set this Alarm Point to 000 to disable. Available
values are:
Ones Place (
0-9
)
Tens Place (
0-9
)
Hundreds Place (
0-9
)
Number of Decimal Places (0, 1, or 2)
Range (
k
on or off – cpm, cps, dpm, Bq; µ, m or none for R/h and Sv/h)
Note:
If the Primary Count Units has changed to a value other than that used
to previously set this Alarm Point, the Alarm Point will be reset to 000.