Model 3001 Multi-Detector Survey Meter
User’s Manual
Section 4
Ludlum Measurements, Inc.
Page 4-11
April 2017
Setup Page 2
Primary RATE/MAX Units and Minimum Display (Default 0 µR/hr) -
Tap ON/ACK to select the value to adjust, and MODE to adjust the value.
Available values are:
Number of Decimal Places (0, 1, or 2)
Minimum Units (See list below.)
cpm kcpm
cps
kcps
dpm
kdpm
Bq
kBq µR/h mR/h R/h µSv/h mSv/h Sv/h
Primary Units RATE/MAX Maximum Display (Default 999 mR/hr) -
Tap ON/ACK to select the value to adjust, and MODE to adjust the value.
Units will be the same as selected earlier with Primary Units. Available values
are:
Ones Place (
0-9
)
Tens Place (
0-9
)
Hundreds Place (
0-9
)
Number of Decimal Places (0, 1, or 2)
Range (
k
on or off – cpm, cps, dpm, Bq; µ, m or none for R/h and Sv/h)
Primary Units RATE/MAX Mode Alarm Point (Default 000) -
Tap
ON/ACK to select the value to adjust, and MODE to adjust the value. Units
will be the same as selected earlier with Primary Units. The ALARM LCD
Segment will be on to indicate an Alarm parameter. Set this Alarm Point to 000
to disable. Available values are:
Ones Place (
0-9
)
Tens Place (
0-9
)
Hundreds Place (
0-9
)
Number of Decimal Places (0, 1, or 2)
Range (
k
on or off – cpm, cps, dpm, Bq; µ, m or none for R/h and Sv/h)
Note:
If the Primary Units has changed to a value other than that used to
previously set this Alarm Point, the Alarm Point will be reset to 000.