Model 702i and 711i
Technical Manual
Section 9
Ludlum Measurements, Inc.
Page 9-12
December 2018
To access any help item, highlight that item and press E
NTER
. If a scroll-bar is shown on the right,
you may wish to scroll through the text using the U
P
and D
OWN
arrow keys. When you are done
reading the help item, press E
NTER
again to return to the menu.
Administrator Menus
The
Administrator
has not only additional menus, but additional menu items on the
Configuration
menu. These allow the system to be customized for a particular application, while still providing
only the most frequently used items to the
User
.
Configuration (Extra Items)
The
Administrator
Configuration
menu is shown in Figure 26. The last three items are identical in
function to those found for the
User
. The additional items are described below.
Figure 26 - Administrator's
Configuration Menu
Dose Rate Alarm Level
This setting determines the dose rate required to set off the dose rate warning alarm.
This is specified in the units given by the
Dose Rate Scale
further down in the menu. The
system will flash the screen and sound an audio alert when this level is reached.
Gamma Rate Alarm Level
This setting determines the gamma count rate in counts per second (cps) that will set off
the gamma warning alarm. Although this has some functional overlap with the dose rate
alarm, there may be cases where a low-energy source with a high count rate would set off