Model 702i and 711i
Technical Manual
Section 8
Ludlum Measurements, Inc.
Page 8-2
December 2018
Figure 15 -
Field Settings Menu with Coarse Calibrate Selected
You will now be prompted to place the
137
Cs source near the detector, as shown in Figure 16. If
you do not have the source nearby, you can ask the system to
Wait
(give you an extra minute) or
Cancel
if you made a mistake.
Figure 16 -
Coarse Calibrate Source Placement Dialog
Once you
Continue
(or allow the countdown to complete) you will see a screen that looks similar to
the one shown in Figure 17. This process may repeat several times as the system updates different
parameters to ensure that your calibration is as close as possible to the target calibration. The target
calibration is shown by the two blue/gray lines near the centers of the peaks. Do not be concerned
if the parameters shown in the message area above the soft keys are different. They are unique to
each detector type and detector.