7.
Connect the voltmeter between the LLD test point (+) on the rear
panel of Slave #1 and ground (-) (the screw head located just to
the right of the LMI logo on the panel). Check that the voltage
here is 100 mVdc ±5%.
8.
Connect the voltmeter between the ULD test point (+) on the rear
panel of Slave #1 and ground (-) (the screw head located just to
the right of the LMI logo on the panel). Check that the voltage is
200 mVdc ±5%.
9.
Observe the count rate for Channel #1. With an output time of
0.50 sec, the count rate should show a value of 333 ±2. With an
output time of 50 msec, the count rate should show a value of 33
±2. Adjust the gain control (GN) clockwise (located on the rear
panel of the slave board) to obtain the 333 ±2 (or 33 ±2) reading,
then adjust the control counter-clockwise to the point where the
counts just cease (less than 10 counts observed).
10.
With the LLD = 100 and ULD = 200 for all channels, observe the
count rate for Channel #1. With an output time of 0.50 sec, the
count rate should show a value of 333 ±2 with the pulser
amplitude set to -15mV. With an output time of 50 msec, the
count rate should show a value of 33 ±2. Adjust the pulser
amplitude toward -20 mV and confirm that the reading drops to 0
(less than 10 counts) at -20 mV ±0.5 mV.
11.
Repeat steps 7 thru 10 above for all remaining channels.
Содержание 4612
Страница 1: ...LUDLUM MODEL 4612 12 CHANNEL COUNTER November 2019 ...
Страница 3: ......
Страница 19: ...Model4612Twelve ChannelCounter TechnicalManual Section2 Ludlum Measurements Inc Page 12 November 2019 ...
Страница 20: ...Model4612Twelve ChannelCounter TechnicalManual Section3 Ludlum Measurements Inc Page 13 November 2019 ...
Страница 35: ...Model4612Twelve ChannelCounter TechnicalManual Section3 Ludlum Measurements Inc Page 28 November 2019 ...
Страница 44: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 37 November 2019 ...
Страница 45: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 38 November 2019 ...
Страница 46: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 39 November 2019 ...
Страница 47: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 40 November 2019 ...
Страница 48: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 41 November 2019 ...
Страница 49: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 42 November 2019 ...
Страница 50: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 43 November 2019 ...
Страница 51: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 44 November 2019 ...
Страница 52: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 45 November 2019 ...
Страница 53: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 46 November 2019 ...
Страница 54: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 47 November 2019 ...
Страница 55: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 48 November 2019 ...
Страница 56: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 49 November 2019 ...