![Ludlum Measurements 4612 Скачать руководство пользователя страница 62](http://html.mh-extra.com/html/ludlum-measurements/4612/4612_manual_1946927062.webp)
Model 4612 Twelve-Channel Counter
Technical Manual
Section 8
Ludlum Measurements, Inc.
Page 55
November 2019
Read Efficiency – REn
Reads the efficiency of the specified channel. The value returned is
formatted as “nn.n.”
Set Efficiency – SEnxx
Sets the efficiency for the specified channel. The efficiency is set using the
format of “nnn.” Do not send the decimal point. The value must be
formatted to three digits and is adjustable from 000 to 999.
Save Slave Flash Parameters – SF
Saves all calibration constants to flash on all slave boards. The calibration
constants will revert to the last saved values if the counter is reset without
saving.
Read High Voltage – RHn
Reads the high voltage set point and read-back value from the specified
channel. The return format is “HVnnnnnnnn.” The first four characters
are the set point; the last 4 characters are the read-back voltage.
Set High Voltage – SHnxxxx
Sets the high voltage set point for the specified channel. The value must be
formatted to four digits and is adjustable from 0000 to 1500.
Read High Voltage Actual Cal – RHACn
Reads the high-voltage actual calibration constant for the specified channel.
The return format is “±n.n.”
Set High Voltage Actual Cal – SHACnxxx
Sets the high voltage actual calibration constant for the specified channel.
The value must be formatted as “±nn.” Do not include the decimal place.
The range is -99 to +99. After setting any calibration constant, the “SF”
command must be sent to save the values into flash.
Read High Voltage Read Back Cal – RHRCn
Reads the high-voltage read-back calibration constant for the specified
channel. The return format is “±n.n.”
Set High Voltage Read Back Cal – SHRCnxxx
Sets the high-voltage read-back calibration constant for the specified
channel. The value must be formatted as “±nn.” Do not include the decimal
Содержание 4612
Страница 1: ...LUDLUM MODEL 4612 12 CHANNEL COUNTER November 2019 ...
Страница 3: ......
Страница 19: ...Model4612Twelve ChannelCounter TechnicalManual Section2 Ludlum Measurements Inc Page 12 November 2019 ...
Страница 20: ...Model4612Twelve ChannelCounter TechnicalManual Section3 Ludlum Measurements Inc Page 13 November 2019 ...
Страница 35: ...Model4612Twelve ChannelCounter TechnicalManual Section3 Ludlum Measurements Inc Page 28 November 2019 ...
Страница 44: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 37 November 2019 ...
Страница 45: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 38 November 2019 ...
Страница 46: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 39 November 2019 ...
Страница 47: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 40 November 2019 ...
Страница 48: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 41 November 2019 ...
Страница 49: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 42 November 2019 ...
Страница 50: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 43 November 2019 ...
Страница 51: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 44 November 2019 ...
Страница 52: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 45 November 2019 ...
Страница 53: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 46 November 2019 ...
Страница 54: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 47 November 2019 ...
Страница 55: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 48 November 2019 ...
Страница 56: ...Model4612Twelve ChannelCounter TechnicalManual Section6 Ludlum Measurements Inc Page 49 November 2019 ...