68
CHAPTER 4. SETUP MODE
• Multiplier with all digits flashing (adjustable from x.xx
µ
to xxx M depending on unit
selected)
Use the UNIT button to select the appropriate parameter, and use the MODE and ON/AUD
buttons to scroll through the parameter options.
Unit 3 - COUNT Units and Minimum Display -
Upon entry into Unit 3 Count Units Mini-
mum Display, one of the following parameters will be selected (flashing).
Note:
Select Unit
first before selecting multiplier.
• Multiplier with all digits flashing (adjustable from x.xx
µ
to xxx M depending on unit
selected)
• Unit (off, d, dpm, Bq, Bq/cm², c, cps, cpm, Rem, Rem/h, R, R/hr, Sv, Sv/h)
Use the UNIT button to select the appropriate parameter, and use the MODE and ON/AUD
buttons to scroll through the parameter options.
Unit 3 - COUNT Alarm 1 -
Upon entry into Unit 3 Count Alarm 1, one of the following pa-
rameters will be selected (flashing).
• Rightmost Digit (0-9)
• Middle Digit (0-9)
• Leftmost digit (0-9)
• Multiplier with all digits flashing (adjustable from x.xx
µ
to xxx M depending on unit
selected)
Use the UNIT button to select the appropriate parameter, and use the MODE and ON/AUD
buttons to scroll through the parameter options.
Unit 3 - COUNT Alarm 2 -
Upon entry into Unit 3 Count Alarm 2, one of the following pa-
rameters will be selected (flashing).
• Rightmost Digit (0-9)
• Middle Digit (0-9)
• Leftmost digit (0-9)
• Multiplier with all digits flashing (adjustable from x.xx
µ
to xxx M depending on unit
selected)
Ludlum Measurements, Inc.
Содержание 3014
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 9: ...CONTENTS 5 A 4 6 AuxComm Setting Auto Mode Interval 96 March 2021 ...
Страница 10: ......
Страница 12: ......
Страница 14: ......
Страница 34: ......
Страница 76: ......
Страница 82: ......
Страница 86: ......
Страница 90: ......
Страница 92: ......
Страница 94: ......