58
CHAPTER 8. ADVANCED FEATURES
The low-battery alarm interval is only active when the device is externally powered. If en-
abled and the battery falls below the low-battery level, a low-battery chime will occur. The
low-battery alarm interval setting defines the number of minutes separating each chime. A
value of 0 will disable the low-battery alarm. This setting can only be configured through
Lumic Calibration software.
8.7 Software Calibration Tools
Lumic Calibration software includes wizards that will assist in calibrating and plateauing
detectors. After configuring the wizard for a specific detector, the wizard will automate much
of the data collection and calculation required for calibration.
8.8 Instrument Calibration
This instrument provides the ability to digitally enter a voltage for both high voltage and
threshold. As such, the instrument itself needs to be calibrated. This happens at the factory
when the instrument is built and should not require re-calibration. However, if calibration is
required (due to board rework, etc.), the calibration settings are available in Lumic Calibra-
tion software.
8.9 DIP Switch Viewing
The current setting of the DIP switches can be viewed from Lumic Calibration software. This
allows the user to check the current DIP switch setting without opening the instrument.
8.10 Other Device Data
The following parameters on the instrument allow recording important device information
within the device:
Firmware Version:
This is a read-only representation of the firmware version. With a version
of F:x.y.zzz, the F:x.y will show up on the device screen during the power-on sequence and
signifies the released version.
Device - Model Name:
This should match the model name on the front face of the instru-
ment.
Device - Serial Number:
This should match the serial number of the instrument.
Detector - Model:
This can store the model of the detector the instrument was calibrated for.
Detector - Serial Number:
This can store the serial number of the detector the instrument
was calibrated for.
Ludlum Measurements, Inc.
Содержание 25024424
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 10: ......
Страница 12: ......
Страница 14: ......
Страница 16: ......
Страница 19: ...C H A P T E R 3 QUICK REFERENCE 3 1 Instrument Front Diagram 15 ...
Страница 20: ...16 CHAPTER 3 QUICK REFERENCE 3 2 Instrument Rear and Sides Diagram Ludlum Measurements Inc ...
Страница 21: ...3 3 INSTRUMENT INTERNAL DIAGRAM 17 3 3 Instrument Internal Diagram June 2022 ...
Страница 22: ......
Страница 26: ......
Страница 34: ......
Страница 38: ......
Страница 58: ......
Страница 64: ......
Страница 70: ......
Страница 72: ......
Страница 76: ......
Страница 88: ......
Страница 90: ......
Страница 91: ......