R-Log – Quick Start 

 

Cod. INSTUM_00953 

Pag. 6 

 

 

3.1.

 

Installing the software on your PC 

 
To configure  your datalogger,  you only have to  install  3DOM  on a PC. However, if this PC is  the 
one  that  will  be  used  for  data  management,  it  is  recommended  to  contextually  install  all  the  other 
software as well along with their licenses of use.  
 

3.1.1.

 

Installation procedure 

 
To install the program, proceed as follows: 
 

 

Introduce 

MW6501 - LSI LASTEM Products

 

DVD

 in the DVD reader of your PC; 

 

Select 

Information-Installation

 in the 

SOFTWARE

 section; 

 

Follow  the  instructions  indicated  under 

Guided  Installation  of  LSI  LASTEM  programs

 

paragraph

.

  

  

3.1.2.

 

3DOM Software 

 
Through  3DOM  software,  you  can  perform  the  instrument  configuration,  change  the  system 
date/time and download stored data by saving them in one or more formats.  
 
At the end of the installation procedure, start 3DOM program from LSI LASTEM programs list. The 
aspect of the main window is as below: 
 
 

 
 

 

.

 

 
 

 
 

Push button selector 

Main menu push button 

Configuration list of 
selected instrument 

Data storage 

supports list 

Configured 

instruments list 

Properties of selected instrument 

Last updated status of 

selected instrument 

Содержание ELR510M

Страница 1: ...Cod MW6165 R Log Quick Start Update 01 25 2016...

Страница 2: ...g a configuration report 14 3 4 Connecting the probes 14 3 4 1 Electrical connection 14 3 5 Displaying measures in fast acquisition mode 16 3 6 Starting the system and displaying measurements 16 4 Ela...

Страница 3: ...DOCUMENTS section 2 To start There are the following models of R Log Model Connector inputs Terminal board inputs Master mode Repeater mode Slave mode Ethernet output ELR510M X X ELR510R X X ELR510S X...

Страница 4: ...R Log Quick Start Cod INSTUM_00953 Pag 4 Legenda ________ cable connection _ _ _ _ _ _ radio connection Fig 1 A type installation Fig 2 B type installation Fig 3 C type installation...

Страница 5: ...each Slave Repeater datalogger with 3DOM software Configuration of the Master datalogger with measurements imported by Slave Repeater devices Generation of a Configuration Report for each datalogger C...

Страница 6: ...Information Installation in the SOFTWARE section Follow the instructions indicated under Guided Installation of LSI LASTEM programs paragraph 3 1 2 3DOM Software Through 3DOM software you can perform...

Страница 7: ...he instrument Insert the instrument in 3DOM Check the instrument s internal clock Create the configuration in 3DOM Send the configuration settings to the instrument For ELR510M ELR510S and ELR510R mod...

Страница 8: ...the screen for about 10 seconds when the window disappears the survey will be automatically started based on the last configuration stored before the instrument shutdown Refer to 3 2 4 if the system...

Страница 9: ...is mandatory and cannot be cancelled In order to have accurate time data the datalogger internal clock should be correct Failing this the clock can be synchronized with that of your computer through...

Страница 10: ...program list Select your instrument serial number in the Instruments panel Select the name of the basic configuration in the Configurations panel user000 by default Press the selected name with the ri...

Страница 11: ...sors registry or you want to connect it to the datalogger in single ended mode in this case refer to the instrument user manual press the New button to add a measure entering all the parameters reques...

Страница 12: ...the instrument The association between input and actuator is fixed and follows the table shown in 3 4 To set an actuation logic proceed as follows Select Logics from the Actuators section Select the...

Страница 13: ...tted via radio to the Master instrument must be entered in the latter as serial measurements To do so proceed as follows Select Measures in the General Parameters section Press Import button to start...

Страница 14: ...later when connecting the probes to the datalogger 3 4 Connecting the probes It is recommended to connect the probes with the instrument turned off 3 4 1 Electrical connection For ELR516M and ELR515...

Страница 15: ...Rx A B C D Current signal from 4 wires sensor powered fixed or switched from E Log 0 V 12V fixed or switched Probe power supply Rx A B C D Rx Current signal from 2 wires probe powered from signal wir...

Страница 16: ...activate the fast acquisition mode proceed as follows Turn the instrument on by simultaneously pressing and buttons Press simultaneously and buttons in the instant values display window Check when pos...

Страница 17: ...me based on the selected settings Basic folder Serial number Prefix _ Serial number _ yyyyMMdd_HHmmss txt Note If the setting Append data on the same file is not selected each time instrument data are...

Страница 18: ...Pag 18 To enable storage select Check to activate data storage control box The list shows the current connection status This can be changed by pressing the Select key that opens the configuration wind...

Страница 19: ...configuration of the programs that use the removed data source and that will continue to use it The timeout for data requests from the database can also be changed To add a new connection select the...

Страница 20: ...ted instrument the Download button is enabled proceed then as follows Select the date from which to start downloading data in case some data were already been downloaded the control proposes the date...

Страница 21: ...Select the identified acquisition with the starting date time of measurements Press the selected acquisition with the right key of your mouse and choose Show Data for the wind direction measurement s...

Страница 22: ...R Log Quick Start Cod INSTUM_00953 Pag 22 To display the chart select Show Chart on the table with the right key of your mouse Enjoy your R Log...

Отзывы: