Thermohygrometers 

 User manual 

 

LSI LASTEM SRL 

INSTUM_04246 

Pag. 13 / 17 

 

5.2

 

Sensing element replacement 

The  sensing  element  is mounted on the stem. The stem replacement  procedure  is similar for all sensors. 
Below we refer to the DMA975 model. 

1.

 

Remove  power  from  the  sensor. 
disconnecting  the  cable  from  the 

Connector  from  DWA  cable

  of  the

 

Case

2.

 

Remove the 

Case

 cover. 

3.

 

Unscrew the wires  from  the 

Terminal 

block

.  

4.

 

Unscrew the 

Case cable gland

5.

 

Unscrew 

Shield  cable  gland

  and  pull 

down the 

Stem

6.

 

Insert the new 

Stem

 in the 

Shield cable 

gland

 until reaching the reference (the 

label); tighten 

Shield cable gland

7.

 

Insert  the  stem  cable  on 

Case  cable 

gland

  and  connect  wires  on  the 

Terminal  block

  as  reported  on  the 

sensor drawing. 

8.

 

Tighten 

Case cable gland

9.

 

Montare il coperchio e collegare il cavo 

Connettore  cavo  DWA

.

 

Mount  the 

case  cover  and  connect  the  cable  to 
the 

Connector for DWA cable

 

 
 

6

 

Accessories / Spare parts 

Code

 

Description

 

SVICA0003

 

ISO9001 type calibration certificate (Temperature) 

SVACA0006

  ISO17025-ACCREDIA type calibration certificate (Temperature) 

SVICA1003

 

ISO9001 type calibration certificate (RH%) 

SVICA1005.1

  ISO17025-ACCREDIA type calibration certificate (RH%) 

DWA505 

Cable L = 5 m 

DWA510 

Cable L = 10 m 

DWA525 

Cable L = 25 m 

DWA526 

Cable L = 50 m 

DWA527 

Cable L = 100 m 

DWA910

 

Extension cable L.10 m for DMA672.5 

DWA925

 

Extension cable L.25 m for DMA672.5 

MG2251 

7 pin free female connector 

ML3015 

Sensitive element (spare part) for EXP815, DMA975-875-867 

DYA049 

Mast mounting device for Ø 45÷65 mm pipe 

EXP301 

Radio signal receiver from EXP815 radio sensor (output compatible with data loggers M/E-Log) 

 

Содержание DMA672.1

Страница 1: ...Environmental monitoring solutions Thermohygrometers User manual ...

Страница 2: ...tained in this manual may be changed without prior notification No part of this manual may be reproduced neither electronically nor mechanically under any circumstance without the prior written permission of LSI LASTEM LSI LASTEM reserves the right to carry out changes to this product without timely updating of this document Copyright 2020 LSI LASTEM All rights reserved ...

Страница 3: ...1 DMA672 x sensors 9 3 2 2 DMA867 DMA875 and DMA975 sensors 10 3 2 3 EXP815 sensor 10 3 3 LSI LASTEM data logger configuration 11 4 Operation check 11 4 1 Calculation of the measurement value based on the sensor analog output 12 5 Maintenance 12 5 1 Cleaning of the anti radiant shield and the porous filter 12 5 2 Sensing element replacement 13 6 Accessories Spare parts 13 7 Disposal 14 8 How to co...

Страница 4: ...uitable for tropical weather in continuous high temperature and RH conditions For outdoor application it should be coupled with a anti radiant shield DMA875 and DMA975 are equipped with a high efficiency natural ventilation anti radiant shield with special black painting on the lower surface of the plates ensures that the sensing element is protected by sun rays for accurate air temperature readin...

Страница 5: ...ower supply 5 24 Vdc Power consumption 2 mA 5 mA Cable and connector L 3 m free wires 8 wires L 1 m male connector for ELU00x enclosures L 3 m male connector for DWA9xx extension cables L 5 m free wires 8 wires detachable Data logger compatibility M Log ELO008 E Log A Log Pluvi ONE uscita RS232 A Log using ALIEM module Pluvi ONE and A Log with ELU00x enclosures Using DWA9nn extension cable M Log E...

Страница 6: ...C 751 table EN 60751 Resolution 0 01 C A M R ELog P1 Response time T90 Typical 4 s 1 m s air flow Long term stability 0 1 C year Relative Humidity Principle Capacitive Measuring range 0 100 Accuracy 1 5 95 DMA672 3 0 5 RH 10 95 Hysteresis 1 Resolution 0 1 A M R ELog P1 0 01 configurable on data logger Response time T90 10 s 1 m s air flow Long term stability 1 year General Information Protection t...

Страница 7: ...ance line of sight 600 m Transmission rate 10 s Battery life 2 years Communication protocol Modbus RTU TTY ASCII Configuration Hyperterminal RS485 protection Galvanic insulation 3 kV UL1577 RS485 speed 1200 115 kbps Power supply Battery AA 3 6 V 10 30 Vac dc Power consumption 10 µW stand by 120 mW in transmission 1 W 3 W Electric protections NO electrically insulated system Tranzorb and Emifilter ...

Страница 8: ...IEC 751 table EN 60751 Response time T90 Typical 4 s 1 m s air flow Long term stability 0 1 C year Relative Humidity Principle Capacitive Measuring range 0 100 Accuracy 1 5 95 Output Programmable RH or Dew Point Resolution 0 1 Response time T90 10 s 1 m s air flow Long term stability 1 year Hysteresis 1 General Information Protection type IP66 Operative temperature 40 80 C Data logger compatibilit...

Страница 9: ...connected with it In order to avoid any damages use this product exclusively according to the instructions herein contained The installation and maintenance procedures must be carried out only by authorized and skilled service personnel Power the instrument in a suitable manner Pay attention and observe the power supplies like indicated for the model in your possession Carry out all connections in...

Страница 10: ...4 Run the cable to the data logger 5 Connect the cable to the data logger as specified in the drawing DISACC supplied with the sensor 3 2 2 DMA867 DMA875 and DMA975 sensors 1 Remove the cover from the sensor box and set the dip switches for 0 20 mA or 4 20 mA output Temperature measurement range Relative Humidity or Dew Point output 2 Fix the cover to the box 3 Fix the DYA049 supporting collar to ...

Страница 11: ...connected wire 3 Read the value displayed on the multimeter and apply the formula shown in 4 1 Voltage output V 1 Set the multimeter to measure voltage signals and scale in V 2 Connect the red test lead of the multimeter to the wire corresponding to the positive signal Sig and the black test lead to the negative signal wire Sig of the sensor it is not necessary to disconnect the sensor from the da...

Страница 12: ...ally at least twice a year it is moreover suggested the replacement of the measure element according to the place of installation in persistent conditions of high humidity pollution dust and chemical substances presence the sensitive element deteriorate faster than the one placed in a location with less adverse conditions It is anyway a good rule to replace the sensitive element at least once ever...

Страница 13: ...on the sensor drawing 8 Tighten Case cable gland 9 Montare il coperchio e collegare il cavo a Connettore cavo DWA Mount the case cover and connect the cable to the Connector for DWA cable 6 Accessories Spare parts Code Description SVICA0003 ISO9001 type calibration certificate Temperature SVACA0006 ISO17025 ACCREDIA type calibration certificate Temperature SVICA1003 ISO9001 type calibration certif...

Страница 14: ...s and disposal lines of this product safeguarding the rights of the consumer Unauthorized disposal of this product will be punished by the law 8 How to contact LSI LASTEM In case of problem contact the technical support of LSI LASTEM sending an e mail to support lsi lastem com or compiling the technical support request module at www lsi lastem com For further information refer to addresses and num...

Страница 15: ...Thermohygrometers User manual LSI LASTEM SRL INSTUM_04246 Pag 15 17 9 CE Conformity Declaration ...

Страница 16: ...Thermohygrometers User manual LSI LASTEM SRL INSTUM_04246 Pag 16 17 ...

Страница 17: ...Thermohygrometers User manual LSI LASTEM SRL INSTUM_04246 Pag 17 17 ...

Отзывы: