LHM - Handmessgerät für Druck, Kra/Gewicht und Drehmoment
LHM - handheld for pressure, force/weight and torque
47
ZERO bu on
Held for 2 seconds, resets the display of the measured value. The display shows
a side. In PEAK mode resets the peak value displayed.
Held for 4 seconds, turns off the ZERO funcon.
PEAK bu on
This key acvates the PEAK mode. Repeated pressing of this bu on cycle-switches
through following modes:
• Normal mode
The actual measured value is displayed.
• PEAK+ mode
The largest value registered is displayed.
• PEAK- mode
The smallest value registered is displayed.
To reset the peak value, press the ZERO key.
The PEAK mode is disabled when the funcon „Time Average“ is acve.
REC bu on
Held for about 4 seconds, this key performs the „Start/Stop“ of a cycle of
data logging.
It is possible to check if the funcons has been accepted through the appearance
of the icon
by side on the display.
If you select the Data Logger Manual mode, by briefly pressing the key a measure
is stored.
BACKLIGHT and REVERSE bu on
Key with two funcons:
Pressed for less than 1 second:
Switches the LCDE backlight on three levels.
Pressed for about 4 seconds:
Switches the display from normal to 180° reverse.
Main Page
On the main page you see the value of the measure according to
the se ngs made in the pages of transducer calibraon.
The value is displayed on a field up to 7 digits including the sign
and decimal point. If the number to be displayed in some
situaons should be greater than 7 digits (including the sign and decimal point) the value is
automacally scaled to one decimal place.
Example 1:
Measured value +100,000
Displayed value: 100,000
The value to be displayed is +100,000. Because the + sign is not displayed, the value is within the
allowable digits.
REC
PEAK
REC
ZERO
REVER
SE