Chapter
8
Maintenance
Repairs
50
09382-03 - 06.05.2021 - LM 2000
Operating inspections
▪
Functionality of LED and display
▪
Measuring accuracy of the temperature sensor
cable and testing for integrity
▪
Accuracy of temperature measurement
▪
Calibration of the temperature test sensor
Electrical safety tests
▪
Operating current
▪
Insulation resistance
Repairs
Repairs may only be carried out by Löwenstein
Medical. For any necessary repairs, please contact
customer service.
(→
Cleaning, disinfection and / or sterilisation
Working with live components!
Risk of injury through electrocution.
▪
Unplug the device from the mains before opening the housing.
▪
Ensure it is not plugged in again without authorisation!
WARNING
Löwenstein Medical has approved the following
methods for cleaning, disinfection or sterilisation.
These methods do not impair the integrity and
function of the humidifier and accessories. The
validation of other methods is the responsibility of the
user.
Cleaning and disinfecting the humidifier
The heating plate on the humidifier can be very hot
Risk of burning
▪
Before cleaning, leave the heating plate to cool.
WARNING
The humidifier and the supplied cables must be
regularly disinfected immediately after use by one
patient and before use by another patient in
accordance with medical instructions.
Follow the disinfectant manufacturer's instructions to
ensure effective disinfection.
Notes on avoiding damage to or
malfunction of the device
▪
Do not subject the humidifier to any sterilisation
process.
▪
Do not use alcohol, solvents or scouring agents.
These substances could damage the device and
cause it to malfunction.