Chapter
6
Operation
Neonate configuration
28
09382-03 - 06.05.2021 - LM 2000
If the humidifier is set to heat both tubes, an alarm
screen will indicate a disconnected heating cable of
the expiration tube.
Pressing (about half a second) the button to pause
the audible alarm confirms the new configuration and
the humidifier continues operation only with the
heated inspiration tube.
If the humidifier is only equipped with the heated
inspiration tube, no alarm is triggered when the
heating cable of the expiration tube is connected and
the humidifier continues operation with the new
configuration (both tubes heated).
When switched on for the first time, the temperature
value at the chamber outlet and near the patient is
set to 34 or 37°C.
In operating mode, the green LED on the device will
flash.
The temperature measured by the temperature
sensor near the patient is displayed in white on the
operating screen below the patient connection
socket.
To access the temperature setting screen
(→
Temperature control screen" p. 29), press the
adjacent button for about half a second.
Press the adjacent button to display the settings
menu.
Humidification can be paused for two minutes by
pressing the [STOP] button for about half a second.
If humidification is paused, the message
“Humidification paused for two minutes” is displayed
at the top of the screen. Simultaneously, the LED on
the device lights up continuously in blue.
The use of this function is recommended when
carrying out procedures that require interruption
of humidification.
You can resume humidification before two minutes
have passed by pressing the [START] button for
about half a second.
The LED on the device lights up continuously in blue
when humidification is paused.