Chapter
6
Operation
Neonate configuration
30
09382-03 - 06.05.2021 - LM 2000
Press the reset button to reset the temperatures to
the default setting. The predefined temperatures at
the outlet of the chamber and near the patient are 34
and 37°C respectively.
The activation of a button by the operator is
confirmed by an audible signal (operation confirmed),
if the button is not pressed long enough, another
audible signal (operation failed) informs the operator
of the failed operation.
For all buttons where no minimum activation time is
specified, activation is immediate when the button is
pressed.
If condensation forms in the inspiration tube, proceed
as follows:
▪
Disconnect the tube and drain the condensed
water into a container. Ensure that it does not
enter the patient's airways during this procedure.
▪
Ensure that the temperature sensor near the
patient is located outside the incubator.
▪
Change the control temperatures (lowering the
temperature at the chamber outlet and/or
increasing the temperature near the patient will
reduce the formation of condensation in the
inspiration tube).
N – List of events
Press the adjacent button, which is located on the
operating screen, to open the list of events.
Opening the screen page will display a list of the
latest events that occurred after the LM 2000 was
switched on.
The events are listed in chronological order. The
most recent event is displayed at the top of the list.
The oldest event is at the bottom of the list.
Pressing the [+] button opens a second page of the
screen with events that occurred even earlier.
Pressing the [-] buttons returns you to the first page
of recent events.
The list of events displays the significant events.
The list of events is reset by switching off the device.
Press the [Cancel] button to return to the operating
screen.