6.7.1
Differential Sensors
Sensors that are classified as differential sensors often utilize a Wheatstone Bridge configuration.
These sensors are essentially a resistive load that use the bridge configuration to detect very small
resistive changes and produce a precise voltage output as a result. Some examples include strain
gauge elements or strain gauge-based sensors, such as some load cells and pressure transducers,
as well as some soil moisture, temperature, and other sensors. For use with the V- Link- 200,
sensors with an impedance of ≥ 120 Ω are recommended.
Calibration in the SensorConnect software for these devices varies depending on the type of sensor
and includes using the shunt calibration for strain gauges. The following diagrams show how to
connect these types of sensors. Note: Full- Bridge is the standard LORD Sensing offering.
Calibration for Half-Bridge, Two Wire Quarter-Bridge, and Three Wire Quarter-Bridge requires the
optional on-board bridge completion.
See Sensor Calibration on page 55
for more information.
Figure 35 - Full Bridge Wiring