Excitation -
Sense -
Signal -
Sense +
Exci
INPUT
REMOTE
Earth
Neutral
Line
AL1
AL1
AL2
AL2
AL3
AL3
AL4
AL4
ALARM RELAY OUTPUTS
POWER
Serial
Output
Cabling should be screened and routed away from noise-carrying cables and
machines to reduce interference. The screen can be connected to power earth if
it is not already grounded at the sensor end.
These connectors are only used when options are fitted
Connections - See also page 23
Neg.
Pos.
E - +
1 2 3 4 5 6
Analogue
Output
Common
Tare
Peak/V
alley
Reset
IMPORTANT!
Set the Lock switches to their ‘ON’ position after setup, to save your settings.
REAR VIEW
Mains Power model
DC Power model
1 2 3 4 5 6 7 8 9 10 11 12 13
14 15 16 17 18 19 20 21 22 23 24 25 26 27
1 2 3 4 5 6
1 2 3 4 5 6
1 2 3 4 5 6
Calibration Method Switch
Alarm Lock Switch
Calibration Lock Switch
Page 5
Hazardous
Area
Hazardous
Area
6 wire Hazardous area schemes. Depending on barriers
chosen, and the number of loadcells used, the excitation
voltage will need to be set appropriately. Our technical
sales team will be happy to advise you. Please note that
we are not in a position to recommend barrier types.
Basic 4 wire system
6 wire system
Safe
Area
Safe
Area
ON
See pages for
different types