Page 26
Troubleshooting help - continued
2. The display value changes when you switch the lockout ON.
This is normal, and can be eliminated. Here is what’s happening ....
At some time in the past, you have tared the display. The tare value is saved in memory.
When you have the lockout switch OFF, the display will always be GROSS, in other words the tare
value will be ignored.
When you have the lockout switch ON, the tare value will be subtracted from the gorss value and the
display will be NETT value.
To clear the tare value to 0, simply reset the meter, when the lockout switch is ON, see below...
This will not affect your calibration.
REAR VIEW
1 2 3 4 5 6 7 8 9 10 11 12 13
ON
1.
Switch 3 = ON
2.
Briefly link terminal 7
to terminal 10 to reset
the Tare memory to 0