Page 9
Features
Calibration Counter / Tamper detector
An internal totaliser counts each calibration. The ‘CAL XX’ value appears for a
second or two after you switch the meter on. The number starts at 00 and can go
up to FF (255 counts). It doesn’t count changes of setpoints, or changes of filter
value or count-by value . It stores the total in non-volatile memory which can’t
be reset, so is useful for keeping track of the meter’s calibration history.
Reset Command
The reset command clears the accumulated total. It may be accessed either
from the front panel or by external contact closure command.
Front Panel key reset command
1) Link terminal 7 to terminal 10
2) Set lockout switch ‘ON’
3) Press Down Arrow key to reset display
Remote contact closure resetting
1) Connect a normally-open contact closure switch between terminals 7 and 10
2) Set lockout switch ‘ON’
Display Rate Command
The ‘displat Rate’ allows you to see the instantaneous flow rate. It may be
accessed either from the front panel or by external contact closure command.
Front Panel key rate display command
1) Link terminal 7 to terminal 9
2) Set lockout switch ‘ON’
3) Press Down Arrow key to see rate
Remote contact closure rate display command
1) Connect a normally-open contact closure switch between terminals 7 and 9
2) Set lockout switch ‘ON’
3) When you make a contact closure between terminals 7 and 9, the display
will show the intantaneous rate value, for as long as the contact is closed.