Clean Earth

Installation  hints  for  best  performance

This section offers several suggestions which will help you get the best performance from
your weighing system.

Loadcells create very small signals which can easily be corrupted by electrical noise. These
steps will ensure you get the best possible performance from your meter.

1.

Use good quality screened signal cable, with 3 twisted pairs. Belden 8777NH,
Belden 9503 and AlphaWire 6010C are good choices, available from many
electrical distributors.

2.

Even if your loadcell is only a 4 wire device, we recommend you use 6 core cable
between the display and the loadcell, and wire as shown in the diagram opposite.
This will ensure that any resistance changes in your cabling, caused by changes
in temperature, will be cancelled.

3.

Each pair should be dedicated to a signal function as shown opposite, for maximum
noise immunity. (One twisted pair for Excitation, one twisted pair for Sense, one
twisted pair for Signal) This will ensure that any electrical noise induced in the
cable is properly cancelled. Mixing functions carelessly amongst the twisted pairs
can actually worsen noise performance.

4.

The cable should be routed away from noisy wiring and devices such as power
feeds from inverters, discharge-lighting cables, welder cabling etc, and should
preferrably be routed in a dedicated low voltage signalling/instrumentation conduit
or cable tray.

5.

Screened cable should be earthed at the display end only.

6.

All wires and screens coming out of the screened cable should be kept as short
as possible to minimise pickup of noise.

7.

If you are using barriers, you should earth your screen as shown below, paying
particular care that you do not earth both ends of any run of of cable.

NB

 If using

galvanic isolation barriers, connect 10K from terminal 2 to terminal 3 and 10K
from terminal 3 to terminal 5 on the display’s Sig. Input connector to maintain the
input signal within the common mode range of the excitation sense circuit.

Barriers

E

A

R

T

H

Connect screen
to earth ONLY
at this end.

Connect screen
to earth ONLY
at this end.

Do not
connect
screen at
this end.

Do not
connect
screen at
this end.

Length of screened cable

Length of screened cable

Loadcell

Hazardous Area          Safe Area

Terminal 7
on logic
input
connector

S

ig

. I

n

p

u

t

L

o

g

i

c

 

I

n

P

o

w

e

r

Display connections

Loadcell connections

Barrier connections

123456

Содержание IL2-L

Страница 1: ...nd general information Caution There is a risk of electrical shock if this instrument is not properly installed Caution Risk of danger Read the whole manual before you install this meter London Electr...

Страница 2: ...2...

Страница 3: ...Calibration Modes 17 Direct Calibration 18 19 Theoretical Calibration 20 21 Sensor Drift correction 22 Logic Input connections Front Buttons 23 Factory defaults 24 CalibrationAudit number 24 Signal F...

Страница 4: ...rs fromelectric shock We designed this meter for Pollution Degree 2 environments only Power supplies to this equipment must have anti surge T fuses rated at 400mA for 230V supply 400mA for 110V supply...

Страница 5: ...hutdown periods you may have to to wait several weeks for an equivalent replacement or we may have no stock at the time you urgently need it You mayalso needto pay extra carriagecharges ifyou wanta fa...

Страница 6: ...240 VAC or 11 30VDC These meters are designed to mount into a protective enclosure which will protect users from contact with power and signal wiring Theseunits mustbe installedfully assembled and mus...

Страница 7: ...of meter U Bracket Bracket screw 1 of 2 Left side of meter case U Bracket Bracket screws U Bracket Bracket screws Slide the meter with gasket into the panel cutout Securing into the panel Re fit the...

Страница 8: ...mple using the 4 buttons Output functions include isolated 16 bit 4 20mA 0 10V 10 to 10V RS232 RS485 and up to 4 alarms with a wide choice of alarm functions Install the meters in a suitable protectiv...

Страница 9: ...sealed IP65 all round and is supplied with the meter in stalled for you to wire to via 3 cable glands It comes complete with SPC4 splashproof cover We also offer a dual version of this enclosure which...

Страница 10: ...5mm2 Tighten screw terminals to 7 0 lb in torque 0 8 Nm torque Cable insulation voltage rating Cables shall have an insulation voltage rating of at least 380V continuous Wire preparation We recommend...

Страница 11: ...nly Link Exc to Sense and link Exc to Sense Set Sig to 5V using 2 x 10K resistors Class II Installation Not used Cable and or zener barrier resistance must not cause the excitation voltage to drop bel...

Страница 12: ...ngst the twisted pairs can actually worsen noise performance 4 The cable should be routed away from noisy wiring and devices such as power feeds frominverters discharge lighting cables welder cabling...

Страница 13: ...13 4 Wire Loadcell 6 Wire Loadcell...

Страница 14: ...ing note to tell you whether it is available in Easy orAdvanced mode How to choose menu mode Lockout Switch must be OFF Press together briefly Set1 Set2 Output Alarms Digit OK Max Min Reset Press brie...

Страница 15: ...witch must be OFF Press together briefly Set1 Set2 Output Alarms Digit OK Max Min Reset Display shows UI Eng Default for User Interface English or UI FrA for User Interface French 1 2 3 Set1 Set2 Outp...

Страница 16: ...ess of the UP button will select a newbrightness level There are 7 brightness levels to choose from Default Full brightness Set1 Set2 Output Alarms Digit OK Max Min Reset Press to accept 1 2 3 Done Di...

Страница 17: ...2 Theoretical Calibration this is when you type in the sensor s theoretical signal level at the bottom and top of itsrange and then type in the value the display should show for each signal level How...

Страница 18: ...et1 Set2 Output Alarms Digit OK Max Min Reset Press OK Apply the highest calibration signal you can achieve ideally 100 of system capacity You can use less but you will get better accuracy with higher...

Страница 19: ...OK Max Min Reset Lockout Switch must be OFF Set1 Set2 Output Alarms Digit OK Max Min Reset Press OK Apply the lowest calibration signal you can achieve ideally 0 of system capacity Set1 Set2 Output Al...

Страница 20: ...ch digit s value as required eg set 20 0000 if your high input is 20mV Set1 Set2 Output Alarms Digit OK Max Min Reset Press to accept Set1 Set2 Output Alarms Digit OK Max Min Reset Display now shows r...

Страница 21: ...ng Use DIGIT to select each digit in turn and the UP or DOWN arrow to increase or decrease each digit s value as re quired eg set 000 000 if your display range is 0 to 100 000 Theoretical Calibration...

Страница 22: ...ings less than 8 least significant counts The corrected value is stored in non volatile memory so that any drift trend will be re ap plied after you have switched off the meter Lockout Switch must be...

Страница 23: ...lines to be galvanically isolated from the input signal To enable the front panel buttons you must connect the logic input for that function to common If you want to have both front panel and remote t...

Страница 24: ...one This feature is available in Easy and Advanced Modes Display shows defs n Defaults no Press the DOWN button to change the display to defs y Defaults Yes if you want to return to default conditions...

Страница 25: ...layed value they will respond at the same rate as the filtered display Lockout Switch must be OFF Press 3 seconds Set1 Set2 Output Alarms Digit OK Max Min Reset Display shows FIL XX X where XX X is th...

Страница 26: ...d up to the nearest 1 no rounding 2 5 10 20 or 50 Lockout Switch must be OFF Press 3 seconds Set1 Set2 Output Alarms Digit OK Max Min Reset Press repeatedly until you see LSt X where X is either 1 2 5...

Страница 27: ...if you wish once you become more familiar with the setup method Lockout Switch must be OFF Press together briefly Set1 Set2 Output Alarms Digit OK Max Min Reset Press repeatedly until you see dLAY XX...

Страница 28: ...Switch must be OFF Press together briefly Set1 Set2 Output Alarms Digit OK Max Min Reset Press OK button briefly and repeatedly until you see boot 0 or boot 1 or boot 2 displayed Set1 Set2 Output Alar...

Страница 29: ...most likely a fault with the display and it should be returned to us for the attention of our QA Manager with a report of what you found for repair c Measure the sense voltage on terminals 2 and 5 an...

Страница 30: ...o dispose of it It must not be thrown away with general rubbish If you are unable to dispose of this item locally you may send it to us for recycling Conditions 1 We will only accept items of our manu...

Страница 31: ...ignals bipolar 4 or 6 wire loadcell up to 40mV 100mV option Input Resistance 10 Megohms Accuracy 0 05 of range Span tempco 25 ppm Degree Celsius Zero Tempco 30 ppm Degree Celsius 5V Excitation voltage...

Страница 32: ...0 Version F0 18 Software released Manual format revised to improve clarity and segregate easy from advanced menu functions Optional outputs now described in their own dedicated manuals DIN Rail mounti...

Страница 33: ...33 Notes...

Страница 34: ...34 Notes...

Страница 35: ...35 Notes...

Страница 36: ...ncludes relay output wiring All signal cabling shall be screened The screen shall only be terminated to the power earth terminal at the meter end of the cable Declared as true and correct for and on b...

Отзывы: