Equipment Specifications
Case Material
Heavy duty welded uPVC
Connectors
Internal detachable Screw Terminal connectors accessed via
compression glands
Environmental
Storage Temperature range -20 to +70C, non condensing.
Allow 30 minutes for display to reach thermal equilibrium.
Operating temperature range 0 to 50C. Internal heater option
available for use in conditions down to -25C
Power
100-240 VAC, 48 VAC, 45 to 60Hz or 11-30 VDC optional
Burden
40VA maximum
Sealing
IP65 all round, provided the display is mounted vertically and that all
cable glands and rear case-closure screws are properly secured.
Input Signals (bipolar)
4 or 6 wire loadcell up to +/-40mV (+/- 100mV option)
Input Resistance
>10 Megohms
Accuracy
+/-0.05% of range
Span tempco
25 ppm/Degree Celsius
Zero Tempco
30 ppm/Degree Celsius
5V Excitation voltage
5VDC nominal rated at 40mA. Must connect to Sense.
10V Excitation voltage
10VDC nominal rated at 120mA. Must connect to Sense.
Excitation Sense
Ratiometric, 4V absolute minimum allowed after line drops
Filtering
/ smoothing
Selectable time constants of 0 to 25 seconds.
A/D conversion
Sigma-Delta 10 conversions per second, 50/60Hz rejection
Resolution 1 in 400 000 max. over full range
Display update rate
10 readings per second.
Display Range (max)
-199999 to +199999, depending on available signal level.
Plug-In Output Options
Analogue, Alarm, Serial, Clock
See separate manuals available from our website
Record of Revisions
17 November 2010
Version F00.19 Software released. Intended for use with 9122-2470 input board,
having switched capacitor input filter feature.
The newly released 9122-2470 input board allows selection of 5V or 10V excitation.
5 January 2011
Version F00.20 Software released. Allows serial data output to derived fron Net or
Gross measurement.
14 February 2011
Version F00.21 Software released. Added permanent segment test mode in bootup
routine options. W arranty increased, terms added.
22 August 2011
Corrected Remote programmer connector details.
Default Input configuration switches
Link S8, S9, S10, S11. Open S2, S3, S4, S12, S13
Excitation voltages
5V DC
Link S11. Open S7 and S12 and S13
10V DC
Link S11 and S7. Open S12 and S13
Excitation Ratiometry
Ratiometric. Link S6. Open S1 (default)
Non ratiometric. Link S1. Open S6
9122-2470
S11 S7
S12 S13
Input Conditioner Board
S2
S8
S3
S9
S10
S4
S6 S1
S5
Содержание Fusion-L 6 digit
Страница 13: ...13 4 Wire Loadcell 6 Wire Loadcell ...