TAMPEREEN AMMATTIKORKEAKOULU
TUTKINTOTYÖ 30(34)
Mikko Huurre
11.5.2 Back section accelerometer
Back section accelerometer is located to the hinge of back part.
Accelerometer is possible to replace as the leg section accelerometer. See
11.5.1
11.5.3 Tilt and lateral sensor
Tilt and lateral sensor is located under the upper body
Replacing the sensor
Drive the table up to the maximum position and on the floor locks
(Lock)
Turn the Stand by switch to the Off position
Remove the column casing (See 8.2.2)
Disconnect the wiring, connector is located next to hydraulic
rotation valve
Undo the screws from the top of rubber cover which is fastened to
the upper cover
Release the sensor
Assemble the accelerometer in reverse order
Be sure that the wiring is same way as before
Check the wiring by moving the tilt and trendelenburg to their
extreme positions. The wiring can not be tight in any positions
11.5.4 Height level sensor
The height level sensor is located in the base of the table. The sensor is
working by using a string to measure the highness. The string is moving
with the table column and it is fastened next to hydraulic pump.