9 TROUBLESHOOTING
Symptom Failure
Action
The cable of the standard hand
control is not connected.
Check the cable connections and
that the cable is intact.
The battery of the auxiliary
(Bluetooth) hand control is empty.
Charge the battery by connecting
the auxiliary hand control to the
operating table with a cable. In
the meanwhile, use the standard
hand control.
Stand-by switch of the table (under
the hatch) is turned OFF
Turn the switch ON
The batteries of the operating table
are empty.
Connect the mains cable to the
charging socket and charge the
batteries.
The hand control has damaged.
Contact service.
In the meanwhile, use the other
hand control
None of the hand control
functions operate and no
leds are visible.
Cable is not connected properly.
Connect the cable as instructed.
Foot pedal do not function.
Foot pedal functions have not been
set up with maintenance software.
Contact service to program the
functions.