background image

 

 

 

Once printed, this document is no longer controlled 
www.cesphn.org.au  

Last updated October 2018 

Page 5 of 7 

 
 

 

 

 

 

5. 

A box (Figure 1) will briefly appear telling you that the Reader has found your LogTag. 

 

 

 

The Reader will read the data from your LogTag and save it to your computer to the default location. 

 

 
Once the data has been saved, LogTag Analyzer will automatically open your report. 

 

 

 

 

 

Clearing, Resetting and re-starting the data logger  

It is best to clear the data after each download so that each weekly download only includes data from 
the previous week’s readings. 
 
Before unplugging the data logger from your computer, click up the top of the software on 

LogTag --> 

Quick Reconfigure.

 This will re-program the logger with the same settings and clear it ready to start 

again.  

 

DON’T FORGET - press and hold the START button again when you are putting it back in 
the fridge.  

 

Press

 

and hold the Start button for about a second. Both the red and green LEDs 

on the front of the logger will flash for a few seconds to show that the logger is 
starting. 
After that, the green OK LED will flash briefly about every 8 seconds to show you 
that the logger is recording. 

 
The data logger is now ready to place immediately back into the fridge. 
 

 

Return the data logger to the fridge asap 

As soon as the report appears, you know that the data has been saved, so it is OK to 

remove the logger from the reader (or the USB port) and return it to the refrigerator.  

Follow the steps below to clear the data logger first and make sure you press START to 

start it recording again).  

You can then come back and analyse your data once the LogTag is back in the fridge. 

Содержание TREX-8

Страница 1: ...818 www cesphn org au Last updated October 2018 Page 1 of 7 User Guide for Operating your LogTag Data Logger Setting up and configuring your Log Tag for the first time 1 Download the software from th...

Страница 2: ...eting old readings to fit them in This way no matter when you download the data you will always record the latest temperatures You should be downloading and clearing your data logger weekly though so...

Страница 3: ...ng red and start flashing green Tick this box to allow this If a staff member ever notices the alert light flashing they should immediately quarantine the fridge until they investigate with a data log...

Страница 4: ...e fridge 3 Connect the data logger to your computer using the Dock or using the USB depending on your model type and Open the LogTag Analyzer software if it doesn t automatically launch for you 4 Down...

Страница 5: ...th the same settings and clear it ready to start again DON T FORGET press and hold the START button again when you are putting it back in the fridge Press and hold the Start button for about a second...

Страница 6: ...ing is OK is to click on the Report Tab at the bottom of the screen At the top left of the Report should be two green ticks showing the Alarm Status for both low and high temperatures If you have a re...

Страница 7: ...hance is to reduce the time the logger is disconnected No health department will have any issue with this To remove the 40 reading from the minimum on the report you can click and drag on the graph to...

Отзывы: