Once printed, this document is no longer controlled
www.cesphn.org.au
Last updated October 2018
Page 3 of 7
After
Because the logger might experience a very short fluctuation in air temperature,
e.g. when someone opens a refrigerator door, we recommend using the first
After
box to set a delay before the Red LED begins flashing. If you set this to 4
consecutive readings, that means, on a five-minute sample rate, the logger will
have to have been out of range for 20 minutes before the Alert light starts flashing.
This will minimise false alarms. Leave the second ‘After’ box clear.
Clear and
reset
The “clear and reset alert” option allows the user to press the “Clear” button on the
logger to stop it flashing red and start flashing green. Tick this box to allow this. If
a staff member ever notices the alert light flashing – they should immediately
quarantine the fridge until they investigate with a data logger download to the
computer.
Passwords
Do not enter a password in the password fields, unless you
specifically
require
the security feature. If you lose or forget the password, there is no way for you to
retrieve it without sending the logger back to the supply company.
Push Button Start
If you selected
Push Button Start
in your LogTag configuration, the logger will now be waiting for you
to start it.
Press and hold the Start (or Start/Mark) button for about a second until both the red and green LEDs
on the front of the logger flash for a few seconds to show that the logger is starting. After that, the
green OK LED will flash briefly about every 8 seconds to show you that the logger is recording.