8.3 Adjusting Parameters for Yeast Cell Counting
<Yeast Counting Parameters on the Protocol Menu>
The adjustable parameters are described as follows:
Parameter
Range
Default
Description
Dilution
Factor
1-10
1.11
The default value for the dilution factor is “1.11”.
To perform the yeast counting, 1 µl of FDA and 1 µl of PI
is added to 18 µl of the yeast sample. The dilution factor
in this case is calculated as 1.11 (20/18). In case 2 µl of
each FDA and PI is added to 16 µl of the cell sample, the
dilution factor is 1.25 (20/16). Likewise, the dilution factor
changes from 1.11(2 µl of the FDA/PI cell viability kit
used), 1.25 (4 µl), 1.42 (6 µl), 1.66 (8 µl), 2.00 (10 µl),
2.50 (12 µl), 3.33 (14 µl), 5.00 (16 µl) to 10.00 (18 µl).
Dilution factor corrects the actual number of cells; thus
the analyzed results are generated for the original
samples.
Minimum
Cell Size
1-90
1
For yeast, the default min value is 1. The unit is 1
micrometer.
Maximum
Cell Size
1-90
20
For yeast, the default max value is 20. The unit is 1
micrometer.
Green
Fluorescence
Threshold
1-10
2
Green and Red Fluorescence Threshold will determine
the level of threshold during the image processing.
Increased threshold will detect less cells by subtracting
Содержание LUNA-FL
Страница 52: ...52 www logosbio com Fluorescence Cell Counting Results...
Страница 63: ...63 www logosbio com 4 8 Printing the Counting Results Please see Section 3 8 to print the counting results...
Страница 93: ...93 www logosbio com 6 6 Printing the Counting Results Please see Section 3 8 to print the counting results...
Страница 114: ...114 www logosbio com Successful Calibration...
Страница 118: ...118 www logosbio com correctly Last try the update again If the problem continues contact your local supplier...