Logos Imaging EPIX Scanner  
User Manual

 

28 

Recommendations on Lighting Conditions 
for Erasing Imaging Plates 

In order to completely erase imaging plates, 
99.5% of the image information must be 
removed. Scanning only removes part of the 
information. To completely erase imaging 
plates, follow the guidelines given below. It is 
important that you measure the light intensity 
with a light meter at the point where the 
imaging plate will be positioned. The intensity 
of light changes dramatically as you move 
away from the source. Erasing the imaging 
plates for periods of time longer than 
suggested does not harm them. A good place 
to store your imaging plates is under the 
erasing light source. 

œ

 

Fluorescent Lights 

A  light box typically gives off between 
1000 and 5000 lux. Measure your light box 
and use the lowest value to determine the 
erasing time. At 1000 lux, an imaging plate 
will be erased in 1 minute. At 2000 lux or 
more, 30 seconds is sufficient. 

œ

 

Incandescent Lighting 

At 1000 lux, an imaging plate will be 
erased in about 2 minutes. At 2000 lux or 
more, erase for 1 minute. 

œ

 

Sunlight 

Full sunlight on a bright day can be as 
strong as 70,000 lux. At this intensity, it 
takes less than 10 seconds to erase imaging 
plates. However, sunlight is too variable, 
and we do not recommend using it. 

If you are unable to directly measure the light 
output from your erasing light source, or do 
not have an erasing light source greater than 
1000 lux, use the practical test, below, to 
determine your erasing time. 

Practical Test 

1.

 

Completely erase an imaging plate by 
subjecting it to direct, intense light for 
about 20 minutes. 

2.

 

Scan the imaging plate. No image should 
appear on the monitor. If an image does 
appear, repeat step 1 with a more intense 
light source. 

3.

 

Place the imaging plate in a sleeve or 
cassette. 

4.

 

Expose the imaging plate using normal 
exposure conditions. If possible, place an 
object, in the path of the x-ray beam. 

5.

 

Turn on your erasing light source. We 
recommend a light box or a minimum l00-
watt incandescent light bulb (see Section 
5.2). 

6.

 

Remove the imaging plate from its sleeve 
or cassette and place it with the blue side 
toward the light for 2 minutes. 

7.

 

Scan the imaging plate. If an image 
appears, the plate was not completely 
erased. Repeat step 6, adding a minute to 
the exposure time. Continue to repeat until 
no image appears on the monitor. Record 
the erasing time that resulted in no image. 
This is the time required to erase an 
unscanned imaging plate. 

8.

 

Repeat steps 1-4. This time, under low 
light conditions, remove the imaging plate 
from the sleeve or cassette and mount it on 
the carousel. Scan the imaging plate. 

9.

 

Remove the scanned imaging plate and 
place it, in the dark, into a light-protected 
area (such as a light-tight drawer. 

10.

 

Now repeat steps 6 and 7, exposing one 
plate (blue side toward light) to the erasing 
light for 2 minutes, the next plate for 3 
minutes, etc., until you find the point 
where no image appears on the monitor. 
This is the time required to erase a scanned 
imaging plate. 

The time required to erase an imaging plate 
will vary depending on the quality and 
intensity of the erasing light source. It should 
be fairly easy to find a light source that will 
erase imaging plates in 2-3 minutes. 

Содержание Digital Imaging System EPIX

Страница 1: ...DIGITAL IMAGING SYSTEM U S Patent Number 5 635 728 User Manual And Installation Guide LOGOS IMAGING EPIXScanner USB LOGOS IMAGING...

Страница 2: ...is a revolutionary new product designed to replace traditional x ray film and film processors TECHNICAL SUPPORT For technical support call you Logos Imaging dealer or Logos Imaging directly SUPPLIES...

Страница 3: ...S 13 Section 5 System Operating Instructions 14 5 1 PREPARING YOUR CURRENT X RAY EQUIPMENT 14 5 2 ERASE THE IMAGING PLATE 14 5 3 TAKING AN X RAY 15 5 4 SCANNING THE IMAGING PLATE 16 5 5 TURNING OFF TH...

Страница 4: ...iographic analysis providing interactive CRT retrieval viewing and processing of stored computed radiographic images The system includes reusable photostimulatable phosphor imaging plates a laser diod...

Страница 5: ...en 100 240 volts AC Imaging plates are toxic The blue colored substance in the imaging plate is toxic Ifpart all of an imaging plate is swallowed contact a physician immediately The physician must rem...

Страница 6: ...sunlight or direct indoor lighting will erase the informationstored on the imaging plate Do not place scanner on or next to a radiator or water source Excessive heat or small amounts of water may dama...

Страница 7: ...Hz Millions of hertz LED light emitting diode mm millimeter cm centimeter kVp peak voltage in thousands of volts IP imaging plates CD ROM compact disc read only CCD charge coupled device TIFF tagged i...

Страница 8: ...8x10 in screen holders 2 8x17 in screen holders 3 8x10 light shields 3 8x17 light shields 1 Logos Imaging software package 1 User manual this manual 1 Software user manual 1 USB cable 6ft 1 Power Cord...

Страница 9: ...canner will not operate with the lid open LED indicator lights with graphic symbols The lights show the current status of the system Green Ready to scan Yellow Scanning Red Error Figure 3 2 Scanner RE...

Страница 10: ...in and 8 x 10 in IMAGING PLATES 8 x 17 in 8 x 10 Plastic mounting strips used to hold imaging plate onto the carousel Active imaging area white color Figure 3 5 CAROUSELS 8 x 10 8 x 17 Top of the caro...

Страница 11: ...er supply auto senses the input voltage Power 100 watts maximum Power Cord Standard line cord provided 3 6 COMPLIANCE TO STANDARDS The EPIX SCANNER Digital Imaging System conforms to the following sta...

Страница 12: ...t countertop large enough to hold the scanner plus provide a working area We suggest a minimum of 18 x 36 in 46 x 92 cm The computer does not need to be on the countertop but must be within the length...

Страница 13: ...relocate your office Inventory the contents of the box and ensure that you have all of the following components If any items are damaged or missing contact your dealer immediately Fig 4 1 Scanner 1 ea...

Страница 14: ...tep 2 Set up the computer and monitor per the manufacturer s recommendations Use an ergonomic setup to minimize repetitive motion injuries Step 3 Turn the power on to the monitor and computer Step 4 C...

Страница 15: ...image information To completely erase the imaging plate expose the front surface blue or white in color to direct intense light for at least 2 minutes An effective way to erase the imaging plate is to...

Страница 16: ...d expose in the usual manner Fig 5 4 In a semi darkened room see Section 4 1 open the screen holder With the white surface pointed away from the carousel insert one side of the imaging plate under the...

Страница 17: ...enough quality erase the imaging plate and start again 5 5 TURNING OFF THE EPIX SYSTEM The EPIX SCANNER Digital Imaging System is designed to be left on continuously The laser is only operational whil...

Страница 18: ...Manual 17 SECTION 6 SOFTWARE OPERATING INSTRUCTIONS Logos Imaging software can capture digital x ray images from the EPIX systems Please refer to the Logos Imaging software user s manual or online hel...

Страница 19: ...is could cause damage to the electronics inside Allow to air dry before plugging in or turning back on Cleaning Imaging Plates Imaging plates should be handled carefully For the best images take care...

Страница 20: ...e cover on the back of the scanner Pry the top of the connector open Fig 7 4 Remove the fuse holder Change both fuses Ensure that new fuses meet the following specifications 250 V 2 amp Time Lag High...

Страница 21: ...n the scanner and computer is loose or defective Reconnect the cable Check for tightness Swap with a known good cable if possible The computer does not recognize that the scanner is connected or the s...

Страница 22: ...ve exposed imaging plates in well lit areas Even in the protective envelope some light penetrates and can partially erase the imaging plates Transfer imaging plates from their protective sleeve to the...

Страница 23: ...Material Items Recycle Comments Plastic Carousel Latches No Dispose of with other non recyclable plastic items Aluminum Housing Optic Module Yes Remove all non aluminum parts before recycling Carousel...

Страница 24: ...t and allow the scanner to come back to room temperature Then follow the guidelines outlined in Section 5 8 2 SHIPMENT The EPIX Digital Imaging System is designed to be shipped in its original shippin...

Страница 25: ...er DHHS Radiation Performance Standards 21 CFR Ch I Subch J EN60825 Class 1 Laser Device Storage and Shipping Conditions 40 to 70 C 40 to 158 F EPIX Imaging Plates Dimensions 8 x 10 in 20 3 cm x 25 4...

Страница 26: ...e with the Logos Imaging written instructions furnished with the unit Please fill out the warranty registration card appendix D and mail to Logos Imaging as instructed EPIX Imaging Plates The EPIX Ima...

Страница 27: ...rtional to the amount of x ray energy the crystal was exposed to The EPIX SCANNER further excites the phosphor crystals to an unstable state by exposing them to a red laser The phosphor crystals then...

Страница 28: ...btain excellent image quality Recommendations on Lighting Conditions for Loading Carousels Fluorescent Lights If loading can be accomplished in less than 60 seconds the fluorescent lighting in the are...

Страница 29: ...Completely erase an imaging plate by subjecting it to direct intense light for about 20 minutes 2 Scan the imaging plate No image should appear on the monitor If an image does appear repeat step 1 wit...

Страница 30: ...d have a minimum of 600 dpi resolution and pseudo randomic dithering Using the standard Windows drivers allows the use of the same printer for all other Windows applications word processing etc The Ep...

Страница 31: ...e continuous improvements to imaging software and you may be interested in some of the upgrades Registering your system with Logos Imaging ensures that you will be promptly notified of any new offerin...

Отзывы: