11 Error messages
E0
⇒
System error.
(Contact service)
E1
⇒
Zero range is higher than 10% of maximum capacity while
switching on the scale.
(example: Scale is loaded while turning on,
problem with plugging in, load cell is damaged)
E2
⇒
Zero range is lower than 10% of maximum capacity while
switching on the scale.
(example: Scale is incorrect sets up, load receptor
is not loaded, problem with turning on, load cell is damaged)
E4
⇒
Zero value is not stable while switching on the scale.
(example:
environment influence, problem with turning on, load cell is damaged)
oF
⇒
Measuring signal out of range
(eg overload, wrong connection,
faulty load cell)
oL
⇒
Overload
(e.g .: Too much load on the scale)
-oL
⇒
Insufficient preload
(e.g.: The preload is smaller than -1/6 of Max capacity)