background image

Interface sheet - Labtech EC-12 R-S - CPET Ergometer

Service documentation

Lode BV
Zernikepark 16
NL 9747 AN Groningen
The Netherlands
t: +31 50 571 28 11
f: +31 50 571 67 46
@: [email protected]
www.lode.nl

INTERFACE SHEET - LABTECH EC-12 R-S - CPET ERGOMETER / VERSION: 1.0 - 2016-01-11

Содержание Labtech EC-12 R-S

Страница 1: ...S CPET Ergometer Service documentation Lode BV Zernikepark 16 NL 9747 AN Groningen The Netherlands t 31 50 571 28 11 f 31 50 571 67 46 ask lode nl www lode nl INTERFACE SHEET LABTECH EC 12 R S CPET E...

Страница 2: ...3 Ergometer setup 4 Setting up the hardware 4 1 Connection overview 4 2 Connecting with the usb cable 4 3 Connecting with the RS232 cable 5 Treadmill or ergometer with 3 5 touch screen unit 6 Treadmi...

Страница 3: ...rm covers All Corival ergometers starting with serialnumber 9609xx 20150001 If your treadmill or ergometer is not covered by this interface form please contact Lode service service lode nl 3 22 INTERF...

Страница 4: ...le to the computer and let Windows install the drivers If your driver is not found you can download the FTDI driver here http lode nl en article ftdi drivers 18 The driver is located under the service...

Страница 5: ...u are using an external blood pressure device Press submit to save the changes The software is now ready for interacing with the ergometer 3 3 Ergometer setup The following protocol needs to be set in...

Страница 6: ...atically You can find the driver on the Lode website under the Services tab Locations For the Treadmill the communication ports are Located on the front side near the power switch For the bicycle ergo...

Страница 7: ...onnect the cable to the Lode ergometer 3 Connect the cable to the external device 4 3 Connecting with the RS232 cable Connecting with the RS232 cable 1 Depending on the external device you may need to...

Страница 8: ...the kind of device the serial number and the connected modules is displayed After initialisation the terminal mode is automatically displayed Swipe with your finger over the screen as indicated with...

Страница 9: ...er The menu Terminal Settings lets you set the communication settings for external communication The menu Service is only intended for service purposes to be executed by factory trainend personel Sele...

Страница 10: ...Select the Communication protocol as desribed in Ergometer setup chapter Select Default for the baudrate 10 22 INTERFACE SHEET LABTECH EC 12 R S CPET ERGOMETER VERSION 1 0 2016 01 11...

Страница 11: ...or USB depending in which cable you have connected to the Lode ergometer Select the button with the house to go back to the main menu 11 22 INTERFACE SHEET LABTECH EC 12 R S CPET ERGOMETER VERSION 1...

Страница 12: ...Select Terminal to set the ergometer in terminal mode Your device is now ready for external control 12 22 INTERFACE SHEET LABTECH EC 12 R S CPET ERGOMETER VERSION 1 0 2016 01 11...

Страница 13: ...the Test window or main menu To access the menu from the Test window you need to swipe the graph area of the touch screen as indicated in the picture Note that for this instruction the graphic view fo...

Страница 14: ...ing the red marked scroll button and select Terminal Select the communication protocol ad described in the Ergometer setup chapter Use the scroll button to view the complete list 14 22 INTERFACE SHEET...

Страница 15: ...elect Default for the baudrate Select RS232 or USB depending on the type of cable that you connected to the Lode ergometer 15 22 INTERFACE SHEET LABTECH EC 12 R S CPET ERGOMETER VERSION 1 0 2016 01 11...

Страница 16: ...ptional Your ergometer can be set to boot up in terminal mode Follow these steps to activate this function Go to the settings menu Select Default Start Menu Select Terminal Switch the ergometer off an...

Страница 17: ...7 Nummerical Control Unit Treadmill only 1 Press and hold key 1 and 2 as shown on Figure 1 for 10 seconds 17 22 INTERFACE SHEET LABTECH EC 12 R S CPET ERGOMETER VERSION 1 0 2016 01 11...

Страница 18: ...uto select The following baudrates can be set Baudrate 0 Autoselect dependent on protocol id 2400 4800 9600 19200 38400 57600 6 Press enter 4 and interface will be displayed you can choose the communi...

Страница 19: ...be downloaded from the Lode website if you are a registered user http lode nl en product getDownload 469 Connect an USB cable from the computer to the USB port of the treadmill or use a RS232 cable fr...

Страница 20: ...ed 3 Select preset Lode Bus Service Protocol 4 Press OK 5 Press Break 6 Switch the device off wait 10 seconds 7 Switch the device on The screen as shown left should show up within 5 seconds after powe...

Страница 21: ...escribed in the Ergometer set up chapter Choose which port you want to use RS232 0 or USB 1 This setting refers to the type of cable that you have connected to the Lode ergometer Select default for th...

Страница 22: ...isable 0 for Global communication timeout Press Y to save the new settings and restart the device Your device is now ready for use with the external controller Note that the device is automatically sw...

Отзывы: