Installation & Start-Up, cont.
3
6
3.2
SPACE & UTILITY REQUIREMENTS
Remove power from all system
components prior to servicing.
Do not defeat lamp system interlocks.
WARNING
LB-11002
EXHAUST AIRFLOW
18 3/8
POWER SUPPLY ENCLOSURE
CURE CHAMBER BASE UNIT
DOOR SWINGS FROM RIGHT TO LEFT
INTERLOCK CABLE
HIGH VOLTAGE LAMP CABLE
POWER INPUT CORD
POWER INPUT CORD
21 1/2
7215
SHUTTER TIMER
CYCLE START
CYCLE STOP
ON
OFF
CHAMBER POWER
P/N 98007
READ OPERATOR'S MANUAL BEFORE USE
prevent user exposure to UV light.
This unit has been designed to
ULTRAVIOLET LIGHT
WARNING
LAMP POWER
CONTROL POWER
OFF
ON
OFF
ON
STANDBY MODE
P/N 984132
MAIN POWER
HIGH MODE
ELAPSED TIME
BEFORE CLEANING, SERVICING OR RELAMPING,
OR LAMP SHIELDING PROVIDED WHEN LAMP IS IN USE.
SUITABLE EYE AND SKIN PROTECTION MUST BE WORN
NOT EXPOSED TO DIRECT OR REFLECTED RADIATION.
MUST BE TAKEN TO INSURE THAT PERSONNEL ARE
WHICH IS HARMFUL TO EYES AND SKIN. GREAT CARE
ULTRAVIOLET IRRADIATORS EMIT RADIATION
ALWAYS TURN THE POWER OFF.
CAUTION
5 INCH EXHAUST VENT
HOSE, MAXIMUM LENGTH
40 FEET, NO INCOMING
POSITIVE PRESSURE
20.00
14.25
10.25
Figure 1. Top View
Figure 2. Front View
Electrical Service Requirements
120 Volts 20 Amp Circuit
High Inrush Magnetic Breaker
120 Volts 15 Amp Circuit
984135 Zeta 7215 & 7216 manl 8/3/00 3:23 PM Page 8