5.2 Positioning
Nitrogen build-up may lead to a lack of oxygen and therefore create the risk of asphyxia.
Always make sure all of the openings of the unit are kept clean and free of obstructions. Keep the
generator away from heat and flames.
The generator must be stored, installed and used only indoor
The generator should be positioned on a flat surface able to withstand a weight greater than 200 kg and
not exposed to vibrations.
Once positioned, lock the wheels using their locking systems to secure the unit in its position and
prevent possible movement
Do not position the generator near naked flames or other sources of heat
Always leave sufficient clearance for the circulation of air around the appliance, as shown below:
A
>= 10 cm
If the unit is positioned under a laboratory desk leave sufficient clearance al the sides and leave
clearance at the rear to permit a good air circulation
Place the unit in a position that allows a comfortable access to the mains plug placed at the back side of
the unit in order to easily disconnect the unit from the supply mains
Do not use the generator in a sealed environment or without suitable ventilation
Do not use the appliance in temperature and humidity conditions outside of the limits specified for
operation (
see 2.3)
These generators must be installed in an environment capable of disposing of the heat produced (see
heat value 2.3)
To perform maintenance operations involving parts inside the generator, the operator needs to access to the
removable side panels and/or the top panel. Place the generator in a space that allows a comfortable access
for maintenance operations.
Operating manual NG(A) CASTORE XL iQ
25