10.3 Alarms
In the event of serious anomalies, nitrogen production stops, the display and the ALARM LED (red) flash
quickly and the buzzer sounds intermittently until the alarm is acknowledged by the user.
To mute the buzzer, simply swipe the touch-screen from the top down, or press the START/STOP button.
#
Message displayed
Cause
Solution
1
Memory data
Error reading the user parameters
Check the user parameters and try
restarting the system,
ONCE ONLY
, if
the problem persists contact service
2
Memory damage
Error reading the configuration
parameters
Try restarting the system,
ONCE
ONLY
, if the problem persists contact
service
3
Inverter
Fault in the device that controls the
motor
Try restarting the system
ONCE
ONLY
, if the problem persists contact
service
4
NOT USED
NOT USED
-
5
Air Pressure Sensor
Damage
Fault reading the internal tank
pressure
Try restarting the system
ONCE
ONLY
, if the problem persists contact
service
6
Head Temp. Too High
Compressor head temperature value
over the allowed limit.
Try restarting the system
ONCE
ONLY
, if the problem persists contact
service
7
Over Pressure
Internal pressure value over the
allowed limit.
Try restarting the system
ONCE
ONLY
, if the problem persists contact
service
8
Air Leaks
The compressor motor has worked for
four hours consecutively without
interruption
-Probables leak in the air circuit
Check for any leaks in the line, if the
problem persists contact service
9
Inverter link fault
Fault communication with internal
inverter
Try restarting the system
ONCE
ONLY
, if the problem persists contact
service
10 Cabinet Temp. Too High
Probable temperature sensor
damaged
Try restarting the system
ONCE
ONLY
, if the problem persists contact
service
11 Dew-point temp High
Dew-point temperature measured
over the threshold
Try restarting the system
ONCE
ONLY
, if the problem persists contact
service
12 Dew-point Sens. damage
Fault reading the dew-point value
Try restarting the system
ONCE
ONLY
, if the problem persists contact
service
13 UEO link fault
Fault communication with UEO
module
Try restarting the system
ONCE
ONLY
, if the problem persists contact
service
14 N2 Press. Too Low
The internal pressure does not reach
the threshold value for more than 1
minute
Check for any leaks in the line, if the
problem persists contact service
Table 11: Alarms
Operating manual NG(A) CASTORE XS iQ
38