9.0 Product Specifications
Power
Power
Power
Power
Power
115 VAC +/-15%, 50/60 Hz
Requirements
Requirements
Requirements
Requirements
Requirements
230 VAC +/-15%, 50/60 Hz
Voltage input selectable via a selector switch located on the I/O PCB.
Fuse: 4A 250 VAC Time Delay
Inputs
Inputs
Inputs
Inputs
Inputs
Flow Switch. All low voltage inputs active low,
i.e. the active state is when the switch is closed. The switch must be capable of
switching 2 mA at +/-15 VDC.
Outputs
Outputs
Outputs
Outputs
Outputs
Pulse Pump A and B, Auxiliary, Alarm.
All low voltage outputs capable of switching 2 mA at = 24 VDC. The pulse output
frequency range is 0-100 SPM. The pulse output is active low. The pulse width is
100 milliseconds in the active low state.
The output type is an opto-isolated NPN transistor open collector configuration.
4-20 mA: 600 ohms maximum
Keypad
Keypad
Keypad
Keypad
Keypad
Three key membrane keypad with tactile response.
Material: Polyester with a hard coat finish.
Actuation Force: 2.6N to 3.3N.
Temperature Input
Temperature Input
Temperature Input
Temperature Input
Temperature Input
Thermistor resistance 10k Ohms at 77° F [25° C]
Circuit Accuracy: +/- 33° F [+/-0.5° C]
Temperature Display: 32° F to 212° F [0° C to 100° C]
Temperature resolution: +/- 1.8° F [+/- 1° C]
Probe Input
Probe Input
Probe Input
Probe Input
Probe Input
Cell constant of 1.5
ESD Protection: 700 Volts
Three (3) sample readings / second : display updated every second
Relays
Relays
Relays
Relays
Relays
Fuse protected electromechanical.
Control Relay (1): 250 VAC, 10amp contact relay
Feed Relay (1): 250 VAC, 10amp contact relay
Biocide Relay (2): 250 VAC, 10amp contact relay
Contact type: Normally open and normally closed contacts (FORM C) Change
over relay.
LCD Display
LCD Display
LCD Display
LCD Display
LCD Display
16-Digit Liquid Crystal Display [LCD] with green backlighting
Operating
Operating
Operating
Operating
Operating
Temperature
Temperature
Temperature
Temperature
Temperature
32° F to 122° F [0° C to 50° C]
Memory Backup
Memory Backup
Memory Backup
Memory Backup
Memory Backup
EEPROM. Data retention of 10 years minimum.
Environmental
Environmental
Environmental
Environmental
Environmental
Printed pc boards conformally coated.
Enclosure: IEC IP65, NEMA 4X - Door closed. IEC IP NEMA - 12X Door open
Mechanical
Mechanical
Mechanical
Mechanical
Mechanical
Two (2) pc boards [three (3) if communication option installed].
Control CPU board: microcontroller and display - low voltage.
Terminal power I/O board: transformer, fuses, terminal blocks, relays.
Option board: 4-20 mA output - low voltage.
(866) 433-6682 • (281) 359-8538 • [email protected] • www.novatech-usa.com