Ethernet Protocol • 146
Gocator 2000 Family
Set Mode
The Set Mode command sets the name of the current system mode.
Command
Field
Type
Description
length
64s
Command size – in bytes.
id
64s
Command identifier (0x1004).
mode[16]
char
Mode name (null-terminated).
Reply
Field
Type
Description
length
64s
Reply size – in bytes.
id
64s
Reply identifier.
status
64s
Reply status.
Get Time
This command retrieves the system clock, in microseconds. All devices in a system are synchronized
with the system clock; this value can be used for diagnostic purposes, or used to synchronize the start
time of the system.
Command
Field
Type
Description
length
64s
Command size – in bytes.
id
64s
Command identifier (0x100A).
Reply
Field
Type
Description
length
64s
Reply size – in bytes.
id
64s
Reply identifier (0x100A).
status
64s
Reply status.
time
64u
Current time, in microseconds.
Get Encoder
This command retrieves the current system encoder value.
Command
Field
Type
Description
length
64s
Command size – in bytes.
id
64s
Command identifier (0x101C).
Reply
Field
Type
Description
length
64s
Reply size – in bytes.
id
64s
Reply identifier (0x101C).
status
64s
Reply status.
encoder
64s
Current encoder value, in ticks.
Содержание Gocator 2000 Family
Страница 1: ...User s Manual Gocator 2000 Family Version 2 2 1 0 Revision A...
Страница 164: ...Specifications 164 Gocator 2000 Family Envelope 61 25 40 5 4 15 82 9 95 7...
Страница 166: ...Specifications 166 Gocator 2000 Family Envelope 33 90 80 2 9 84 3 99 7 10 6...
Страница 168: ...Specifications 168 Gocator 2000 Family Envelope 190 210 33 141 3 155 7 4 11...
Страница 170: ...Specifications 170 Gocator 2000 Family Envelope 33 300 400 214 9 229 2 10 3 3 2...
Страница 172: ...Specifications 172 Gocator 2000 Family Envelope 33 400 500 216 2 230 6 3 9 11...
Страница 174: ...Specifications 174 Gocator 2000 Family Envelope 33 350 800 A 3 7 10 8 215 8 230 3 DETAIL A SCALE 1 2...