The result of each measurement (blood pressure and heart rate) is automatically stored in the memory of device.
DATA IN MEMORY WILL BE SAVED EVEN IN THE DEVICE WITHOUT BATTERIES. IN ORDER TO DELETE SAVED DATA FROM THE MEMORY
DEVICE,FOLLOW THE STEPS DESCRIBED IN THE “MEMORY FUNCTION”.
If the power is not switched off and the appliance is not used for 3 minutes, then switches off.
AUTOMATIC INFLATING
If the initial pressure level (190 mm Hg.) is insufficient or there is movement of the hand during the measurement, the unit will stop measuring and
inflate the cuff to the next higher level of pumping. The device has 4 fixed inflation level of the cuffs: 190, 230, 270, 300 mm Hg.
Automatic inflating of the cuff is repeated until the measurement is completed successfully. This is not a malfunction.
FORCED RELEASE OF PRESSURE FROM THE CUFF
If during the measurement you need to relieve quickly the pressure in the cuff - touch
. The device release quickly the air from the cuff and off.
MEMORY FUNCTION
1. The result of each measurement (blood pressure and heart rate) is automatically stored in the memory of device.
MEASUREMENT RESULT WILL NOT BE SAVED IN CASE OF THE NOTICE OF ERROR.
2. The memory device can save up to 90 measurements and average of the last 3. When the number of measurements exceeds
90, the oldest data is automatically replaced with the following data measurements.
3. You can view the contents of the memory device by touching the key M. The first time you touching the key M, the
symbol “A” on the screen briefly displays and the average value of all readings stored in memory (Fig. 5). When you touch
the key M again, the index “1” (the number of memory cells) displays briefly on the screen , then the result will be the last
measurement (Fig. 6).