If irregular pulse rhythm is detected during measurement, symbol of arrhythmia
“
”, will appear upon measurement end. Appearance of arrhythmia indicator
may also be caused by body movement during measurement. During
periodical appearance of this indication apply to You attending doctor.
Apart from numerical value of pressure, result is also
displayed on scale WHO (Fig. 14). Scale WHO – three-
color scale of classifi cation of received value of arterial
pressure, according to recommendation of World Health
Organization. The scale is available from the left.
6. Press to switch off .
TO OBTAIN THE ACCURATE RESULT MAKE INTERVAL
BETWEEN MEASUREMENTS TO RESTORE THE BLOOD
CIRCULATION. WAIT FOR AT LEAST 3 MINUTES BEFORE
MAKING A NEW MEASUREMENT.
THE DATA WILL BE KEPT IN THE MEMORY EVEN IF THE DEVICE IS STORED
WITHOUT BATTERIES. TO DELETE ALL VALUES STORED IN THE MEMORY YOU
SHOULD MAKE ACTIONS DESCRIBED IN “MEMORY FUNCTION”.
If no actions are done in mode of date and time setting for more than 3 minute,
the Device is independently changed over to mode of date and time indication.
AUTOMATIC RE-INFLATION
When during the fi rst blood pressure measurement the cuff infl ation to a
level of 190 mmHg is not suffi
cient or you move your arm the Device stops
measurement and re-infl ates the cuff to the higher level. The Device has 4
fi xed levels of the arm cuff infl ation: 190, 230, 270 and 300 mmHg.
The automatic re-infl ation is repeated until the measurement is completed
successfully. This is not a defect.
FORCED DEFLATION FROM A CUFF
For rapid air release from of the arm cuff dur-
ing arm cuff infl ation or during a measurement
(slow defl ation) press the Button. The device
will quickly release all air from cuff and will
switch off .
MEMORY FUNCTION
1. Result of each measurement (pressure, pulse, time and date of measurement)
may be kept in the device memory. For this purpose, after measurement, within
not more than 3 minutes, memory M1 or M2 shall be selected for memory
storage.
IF THE NOTICE ON ERROR APPEARS THE MEASUREMENT RESULT WILL
NOT BE STORED.
M
D
Fig.15
Fig.16