6. If the arm has a conic form, the cuff should be
put on with a spiral movement as shown in the
fi gure.
Fig.11
7. If the rolled-up sleeve squeezes the arm
interfering with free blood flow the Device
may give inaccurate fi gures not corresponding
to your actual blood pressure.
Fig.12
MEASUREMENT PROCEDURE
1. Insert the Air Tube Plug into the Cuff Jack.
Make 3-5 deep inhales and exhales before taking a measurement and relax.
Do not move, do not speak and do not toughen your arm.
2. Press .
3. All symbols will appear on the display screen for a short time (fi g. 13), two
short sound signals will be given and the Unit will infl ate automatically the
air into the cuff .
At fi rst the infl ation will stop at the level of 190
mmHg (Fig. 14).
4. After reaching the level of 190 mmHg the cuff
will gradually defl ate. The fi gures on the screen
will count back. The pulse symbol “ ” will start
fl ickering.
AS THE BLOOD PRESSURE AND PULSE ARE
MEASURED DURING AIR DEFLATION FROM THE
ARM CUFF KEEP QUIET AND DO NOT MOVE
YOUR ARM AND DO NOT TOUGHEN YOUR ARM
MUSCLES.
5. When the measurement is complete the sound
signal is given, the arm cuff completely defl ates
and your measurement results flash on the
screen (Fig. 15).
If irregular pulse rhythm is detected during
measurement, symbol of arrhythmia «
»,
will appear upon measurement end (for LD23,
LD23A). Appearance of arrhythmia indicator may also be caused by body
movement during measurement. During periodical appearance of this indication
apply to You attending doctor (for LD23, LD23A).
Fig.14
Fig.13
Fig.15