10
ENG
6. Press the to switch off the Unit. For taking a new measurement repeat all
steps described in this paragraph.
TO OBTAIN THE ACCURATE RESULT MAKE INTER VAL BE T WEEN
MEASUREMENTS TO RESTORE THE BLOOD CIRCULATION. WAIT FOR AT LEAST
3 MINUTES BEFORE MAKING A NEW MEASUREMENT.
If the Unit is ON and is unused for 3 minutes it will be switched off
automatically.
AUTOMATIC RE-INFLATION
When during the fi rst blood pressure measurement the cuff infl ation to a
level of 190 mmHg is not suffi
cient or you move your arm the Unit stops
measurement and re-infl ates the cuff to the higher level. The Unit has 4 fi xed
levels of the arm cuff infl ation: 190, 230, 270 and 300 mmHg.
The automatic re-infl ation is repeated until the measurement is completed
successfully. This is not a defect.
FORCED DEFLATION FROM A CUFF
For rapid air release from of the arm cuff during arm cuff infl ation or during a
measurement (slow defl ation) press the . The arm cuff quickly completely
defl ates
and the Device is independently switched off .
INFORMATION ABOUT ERRORS
Indication
Likely cause
Methods of correction
The arm cuff is applied incorrectly
or the air tube plus is inserted not
tightly enough.
Measurements cannot be made
due to hand movement or talking
during measurements.
At such disorders as deep vascular
sclerosis, weak pulse wave and
also in patients with the prominent
distortions of cardiac rhythm it
may be difficult to measure the
blood pressure accurately.
Be sure that the arm cuff is
applied correctly and the plug
is inserted tightly. Repeat
the whole measurement
procedure.
Repeat the measurement
following strictly the recom-
mendations of this Manual.
In such cases consult a
cer tified doc tor about
application of the electronic
unit.
Batteries are discharged.
Replace the batteries for new
ones.