background image

                                                                                   

Shenzhen Linshang Technology Co., Ltd.

 

3/12/2022      Page 7 of 8 

2)    Press the probe against the tested object surface vertically. Keep the probe stable, do not tilt or shake. 

Measurement value will appear on the display with buzzer alerts.   

3)    To continue the measurement, you can lift off the probe away from the object to be measured, and 

then follow step 2) again. 

4)    When the device identifies the ferrous putty, it will turn on the red backlight. The buzzer alerts for two 

times.  And  the  interface  of  the  device  will  prompt:  “Ferrous  putty”.  When  the  device  identifies  the 
galvanized iron, it will turn on the green backlight. The substrate is displayed as “FeZn”.   

5. Check measurement records 

In measurement mode, short press the button to check historical data. The gauge stores 9 sets of data. 
When more than 9 sets of data are stored, the earliest recorded value is automatically deleted, and “No.1” 
is the last test data. Recorded data will not loss after powering off. 

V. Precaution 

1.  The device must be zero adjusted respectively with the iron-base and aluminum-base zero adjustment 

plates. Otherwise, there may be abnormal identifications of the ferrous putty and galvanized iron substrate.     

2.  It is strictly forbidden to slide the probe on the surface of the car body; otherwise both the car paint and 

device will be damaged.   

3.  Please ensure that the paint surface of the car body is clean. Dust, dirt and other foreign objects on the 

surface will affect the measurement accuracy.   

4.  When the device displays “Low Battery”, batteries should be replaced. 

VI. Packing list 

No. 

Description 

Quantity 

Unit 

Coating Thickness Gauge 

Set 

Fe zero-adjustment plate 

pcs 

NFe zero-adjustment plate 

pcs 

Standard film   

pcs 

User manual 

pcs 

Drawstring bag 

pcs 

Certificate / Warranty card 

pcs 

Содержание LS236

Страница 1: ...Shenzhen Linshang Technology Co Ltd 3 12 2022 Page 1 of 8 LS236 Coating Thickness Gauge User Manual V2 0 Please read this manual carefully before using and reserve it for reference ...

Страница 2: ...isible under strong light The top display uses OLED OLED ensures normal display in a low temperature environment of 40 C It also features 3 color backlight display When the device identifies the ferrous putty the backlight turns red In case of the galvanized iron substrate the backlight turns green And under other situations the backlight is white Standards for the product DIN EN ISO 2808 Paints a...

Страница 3: ...ent angles 5 Using LCD and OLED as display devices simultaneously LCD ensures that the test results are clearly visible under strong light and OLED ensures normal display in a low temperature environment of 40 C 6 The device can identify the ferrous putty as well as the galvanized iron substrate It also features 3 color backlight display 7 Wear proof ruby probe tip for long term use 8 Dual use pro...

Страница 4: ...ess than 5 s to confirm the settings long press the button for 5 s the gauge exits the setting and shuts down and the setting is invalid Language settings The instrument is available in Chinese English Russian Turkish Setting method In the main setting interface short press power button to select Language option long press the button for 3 seconds to confirm the selection short press the button ag...

Страница 5: ...rm and back to the main setting interface Main setting interface Unit selection interface Exit In the main setting interface short press power button to select Exit long press the button for 3 seconds to confirm and enter the measurement interface 3 Zero adjustment A zero adjustment is required when using the gauge for the first time after inserting new batteries working with different materials o...

Страница 6: ... 3 4 The zero adjustment completed when there is the buzzer sound again and the LCD screen displays 0 0 3 5 After the zero adjustment is completed place the standard film on the plate substrate if the measurement value is stable and deviation from standard value within 3μm the gauge can be used normally Note After the zero adjustment completed when repeating measurement on the same spot the readin...

Страница 7: ... than 9 sets of data are stored the earliest recorded value is automatically deleted and No 1 is the last test data Recorded data will not loss after powering off V Precaution 1 The device must be zero adjusted respectively with the iron base and aluminum base zero adjustment plates Otherwise there may be abnormal identifications of the ferrous putty and galvanized iron substrate 2 It is strictly ...

Страница 8: ...ally please send the whole gauge to our company for maintenance 2 Provide users with spare parts and lifelong maintenance services 3 Provide the users with the gauge calibration service 4 Free technical support for long term Manufacturer Shenzhen Linshang Technology Co Ltd Website www linshangtech com Service hotline 86 755 86263411 Email sales21 linshangtech com ...

Отзывы: