Shenzhen Linshang Technology Co., Ltd.
3/12/2022 Page 5 of 8
Setting method: In the main setting interface, select "mode" option by short pressing power button, and
long press the button for 3 seconds to confirm the selection. Shortly press the button to select the
measurement mode. And long press the button for 3s to confirm the selection and back to the main
setting interface.
Unit setting
The device can be set to the metric or imperial unit, and the factory default is metric.
Setting method: In the main setting interface, shortly press the button to select “Unit”. Long press the
button for 3s to confirm and then enter the unit selection interface. Shortly press the button to select your
unit and long press the button for 3s to confirm and back to the main setting interface.
Main setting interface Unit selection interface
Exit
In the main setting interface, short press power button to select "Exit", long press the button for 3
seconds to confirm and enter the measurement interface
3. Zero adjustment
A zero-adjustment is required when using the gauge for the first time, after inserting new batteries, working
with different materials or ambient temperature changes. We strictly recommend carrying out the reference
check on the uncoated original substrate, due to the difference of magnetic and conductive properties of
the material, some measurement deviation will be caused. If this is not possible, please use the zero
reference plates supplied by Linshang technology, there are Fe plate and NFe plate, please choose
correctly according to the measuring materials.
3.1 Measuring the plate or uncoated original substrate, a measured value will be displayed on the gauge,