background image

                                                                                   

Shenzhen Linshang Technology Co., Ltd.

 

3/12/2022      Page 5 of 8 

Setting method: In the main setting interface, select "mode" option by short pressing power button, and 
long press the button for 3 seconds to confirm the selection. Shortly press the button to select the 
measurement mode. And long press the button for 3s to confirm the selection and back to the main 
setting interface.   

 

Unit setting 

The device can be set to the metric or imperial unit, and the factory default is metric. 

Setting method: In the main setting interface, shortly press the button to select “Unit”. Long press the 
button for 3s to confirm and then enter the unit selection interface. Shortly press the button to select your 
unit and long press the button for 3s to confirm and back to the main setting interface. 

 

 

                          Main setting interface                                      Unit selection interface 

Exit 

In the main setting interface, short press power button to select "Exit", long press the button for 3 
seconds to confirm and enter the measurement interface 

 

3. Zero adjustment 

A zero-adjustment is required when using the gauge for the first time, after inserting new batteries, working 
with different materials or ambient temperature changes. We strictly recommend carrying out the reference 
check on the uncoated original substrate, due to the difference of magnetic and conductive properties of 
the  material,  some  measurement  deviation  will  be  caused.  If  this  is  not  possible,  please  use  the  zero 
reference  plates  supplied  by  Linshang  technology,  there  are  Fe  plate  and  NFe  plate,  please  choose 
correctly according to the measuring materials. 

3.1 Measuring the plate or uncoated original substrate, a measured value will be displayed on the gauge, 

Содержание LS236

Страница 1: ...Shenzhen Linshang Technology Co Ltd 3 12 2022 Page 1 of 8 LS236 Coating Thickness Gauge User Manual V2 0 Please read this manual carefully before using and reserve it for reference ...

Страница 2: ...isible under strong light The top display uses OLED OLED ensures normal display in a low temperature environment of 40 C It also features 3 color backlight display When the device identifies the ferrous putty the backlight turns red In case of the galvanized iron substrate the backlight turns green And under other situations the backlight is white Standards for the product DIN EN ISO 2808 Paints a...

Страница 3: ...ent angles 5 Using LCD and OLED as display devices simultaneously LCD ensures that the test results are clearly visible under strong light and OLED ensures normal display in a low temperature environment of 40 C 6 The device can identify the ferrous putty as well as the galvanized iron substrate It also features 3 color backlight display 7 Wear proof ruby probe tip for long term use 8 Dual use pro...

Страница 4: ...ess than 5 s to confirm the settings long press the button for 5 s the gauge exits the setting and shuts down and the setting is invalid Language settings The instrument is available in Chinese English Russian Turkish Setting method In the main setting interface short press power button to select Language option long press the button for 3 seconds to confirm the selection short press the button ag...

Страница 5: ...rm and back to the main setting interface Main setting interface Unit selection interface Exit In the main setting interface short press power button to select Exit long press the button for 3 seconds to confirm and enter the measurement interface 3 Zero adjustment A zero adjustment is required when using the gauge for the first time after inserting new batteries working with different materials o...

Страница 6: ... 3 4 The zero adjustment completed when there is the buzzer sound again and the LCD screen displays 0 0 3 5 After the zero adjustment is completed place the standard film on the plate substrate if the measurement value is stable and deviation from standard value within 3μm the gauge can be used normally Note After the zero adjustment completed when repeating measurement on the same spot the readin...

Страница 7: ... than 9 sets of data are stored the earliest recorded value is automatically deleted and No 1 is the last test data Recorded data will not loss after powering off V Precaution 1 The device must be zero adjusted respectively with the iron base and aluminum base zero adjustment plates Otherwise there may be abnormal identifications of the ferrous putty and galvanized iron substrate 2 It is strictly ...

Страница 8: ...ally please send the whole gauge to our company for maintenance 2 Provide users with spare parts and lifelong maintenance services 3 Provide the users with the gauge calibration service 4 Free technical support for long term Manufacturer Shenzhen Linshang Technology Co Ltd Website www linshangtech com Service hotline 86 755 86263411 Email sales21 linshangtech com ...

Отзывы: